Mosaic3
42
Version 3.5 rev 8 May 2019
INSTALLATION
4.5 m
osaic
c
abLing
All device interconnection cables will be installed by the Andor Installation Technician and should not be
removed by the user.
The External Trigger and DMD Active Input and Output use SMA connectors. If connecting a new cable carefully align
the cable’s central pin with the coupler on the Mosaic head then gently insert the cable into the coupler. While holding
the cable in place turn the locking collar clockwise until its locked finger tight, do not use a tool to tighten further.
Removal is the reverse of this process.
4.6 m
osaic
o
pticaL
f
ibre
This is a non-user serviceable item and should not be disconnected from the Mosaic head or the Laser Source. If there
is a problem, please contact Andor technical support (see
Section 1.1 Help & Technical Support
).
4.7 s
oftWare
i
nstaLLation
If the device is supplied with Andor’s iQ Software then the Installation Technician will ensure that the software is installed
and operating correctly. For all other software programs please see the applicable User Guide for details. Future
upgrades can be installed by the customer. The latest Andor software releases are available through MyAndor:
http://my.andor.com/login.aspx.
4.8 r
emote
i
nterLock
All laser based Mosaic systems are supplied with a safety interlock kit which will be installed and tested by the Andor
Installation Technician.
If further external interlocks, e.g. doors, need added, the modular nature of the interlock cable assembly allows for
further interlocks to be added. Please contact Andor Customer Support to arrange this.
4.9 s
ystem
c
omponents
and
a
ccessories
Only components and accessories specified and supplied by Andor are suitable for use with the Mosaic. Please contact
your local Andor Representative for assistance with any accessory requirements.