5.Advanced User Guidebook
69
Figure 5-67 Measure Menu
The "Automatic Measurements" menu is described as the following table:
Function Menu
Setting
Description
Add
Type
Press F1 ,show the measure types
Source
CH1
Select the source
CH2
Show all
Show all the measures on the screen
Add
Add the selected measure types (shown
at the left bottom, you could only add 8
types at most)
Remove
Remove all
Remove all the measures
Type
Turn the
M
knob, select the types need
to be deleted.
Remove
Remove the chosen measure type
Measure
The measured values can be detected on each channel simultaneously. Only if the
waveform channel is in the ON state, the measurement can be performed. The automatic
measurement cannot be performed in the following situation: 1) On the saved waveform.
2) On the mathematical waveform. 3) On the XY format. 4) On the Scan format.
Measure the period, the frequency of the CH1 and the mean, the peak-to-peak voltage of
the CH2, following the steps below:
1.
Press the
Measure
button to show the automatic measurement function menu.
2.
Press the
H1
button to display the
Add
menu.
3.
Press the
F2
button and choose
CH1
as the source.
4.
Press the
F1
button, the type items will display at the left of screen, and turn the
M
knob to choose
Period
.
5.
Press the
F4
button, the period options added completes.