EN
Owandy Visteo – User manual
Page 14/30
In the configuration window:
◄
Check the “USB” model.
◄
Select the “Visteo” kit.
◄
Set the activation time of the box
(default 5min).
◄
Set the inlay and size of the date & time
and exposure parameters in the
acquired image.
◄
Select the image treatment (*).
◄
Configure the Owandy software (**).
(*) When the “Film alike” option is activated, the contrast depends on the exposure time. Adjust the X-ray dose on
the generator to obtain a good image.
(*) When you select the option “Auto contrast” the contrast is constant. Exposure errors are corrected
automatically, which reveals noise in badly exposed images.
In both cases, the exposure bar (blue/green/red) helps to find the correct exposure of the images.
(**) This option appears only when the sensor is used from certain Owandy software programs; it allows the
change of the sharpness setting or the high resolution for each acquired image.
5. Click on “OK” to confirm your choice.
6. Then click on the “Save”
button (QuickVision 3) or on the “Validate” button (QuickVision 4)
to validate the settings.
The use of the kit is identical to the use of the Owandy XIO StandAlone software described below.