E-216-5N-E
33
If a password is set, make a note
not to forget it.
If you forget the password,
please contact business office or
service center of our company.
[INPUT PASSWORD] screen shown right
is displayed.
Carry out the setting using a numeric
keypad and CLR button. If you touch SET
button, the password will be set. If you
touch CANCEL button, the password will
be cancelled.
In the cases (1) to (5) shown above
If the password is correct, the screen will shift to each screen.
NOTE:
[INPUT PASSWORD] screen
Password function
●
Setting of password (at initial setting)
●
Timing when the input of password is required
If a password is set, the input of password is required in the timings (1) to (5) shown below. For the current
value of password, "* (asterisk)" is displayed for the number of setting digit.
(1) When you touch each data field on [PARAMETER] screen (when you want to change a parameter)
⇒
If you input the password once, re-input of password will not be required until quitting [PARAMETER] screen.
(2) When you touch "1.PASSWORD" (when you want to change the password)
(3) When you touch "2.IN/OUT.CHECK"
(4) When you touch "3.TOUCH.ADJUST"
(5) When you touch "4.INITIALIZE"
Touch "4.SYSTEM" on [MENU] screen to shift to [SYSTEM] screen.
Next, touch "1.PASSWORD".
[INPUT PASSWORD] screen will be displayed. Input "111111 (default)".
After the password is input, touch SET button to shift to [PASSWORD SETTING] screen.
Change the password using a numeric keypad and CLR button.
・
Max. 6-digit arbitrary number can be set.
* It is also possible to delete the password by clearing the number and touch SET button.
If you touch SET button twice within one second, the changed password will be memorized and the screen
will return to [SYSTEM] screen. If you touch CANCEL button, the change will be cancelled and the screen
will return to [SYSTEM] screen.
Then, touch "MENU" and "5. BATCH PROCESSING" buttons and return to the measurement screen.
Fig. 9.3 How to set a password
"1.PASS
WORD"
[SYSTEM] screen
[PASSWORD SETTING] screen
"SET"