L-737-3-E
7
9.15.4 Totalizer 1 reset (Reset Total 1)
The transmitter is provided with counters (Counters and ) and totalizers (Totalizers and ) which
work in synchronism with pulse output. When an input arrives across status input terminals, counter
and totalizer are reset to 0.
In this connection, counter and totalizer respond to pulse output while counter and totalizer
respond to pulse . (For detail of the function, see .0 Reset Functions.)
() Operating procedure
・
Select "Reset D/T " in selecting the status input function assignment.
(See .. Transition chart () [No. to -].)
・
Set the input mode of status input. According to the settings, the following input mode will be
enabled:
"Short Active" setting: the function is activated 0.3 seconds after the condition of the circuit
between status input terminals is changed from opened to shorted.
"Open Active" setting: the function is activated 0.3 seconds after the condition of the circuit
between status input terminals is changed from shorted to opened.
(See .. Transition chart () [No. to -])
The function becomes available by setting the above.
9.15.5 Totalizer 2 reset (Reset Total 2)
The transmitter is provided with counters (Counters and ) and totalizers (Totalizers and ) which
work in synchronism with pulse output. When an input arrives across status input terminals, counter
and totalizer are reset to 0.
In this connection, counter and totalizer respond to pulse output while counter and totalizer
respond to pulse . (For detail of the function, see .0 Reset Function.)
() Operating procedure
・
Select "Reset C/T " in selecting the status input function assignment.
(See .. Transition chart () [No. to -].)
・
Set the input mode of status input. According to the settings, the following input mode will be
enabled:
"Short Active" setting: the function is activated 0.3 seconds after the condition of the circuit
between status input terminals is changed from opened to shorted.
"Open Active" setting: the function is activated 0.3 seconds after the condition of the circuit
between status input terminals is changed from shorted to opened.
(See .. Transition chart () [No. to -])
The function becomes available by setting the above.