Setup
Default
3---
Hardware
Runs a check for possible problems in the probe.
For details, see .. Probe check.
CAUTION: Be very careful as this check can stop outputs under certain circumstances.
See ..
—
3---
Drive coil Check
Runs a check for an "open" in the drive coil.
For details, see .. Drive coil check.
CAUTION: Be very careful as this check can stop outputs under certain circumstances.
See ..
—
3---3
Txr Condition
Runs a check for possible problems in the transmitter.
For details, see ..3 Transmitter check.
See ..3
—
3---- Back Light
Tests backlight for possible problems by forcibly turning it on and off. While test continues, the LCD remains dead.
Performance test consist of WHT backlight lit (3 sec), ORG backlight lit (3 sec) and backlight unlit (3 sec) conducted
twice one after the other (total of sec) and ends automatically. After the test, the LCD restores normal illumination.
See .
—
3---- LED
Tests LEDs for possible problems by forcibly turning them on and off. While test continues, the LCDs remain
dead. Performance test consist of RED LED lit (. sec) and GRN LED lit (. sec) conducted times
alternately (total of sec) and ends automatically. After the test, the LCDs restore normal illumination.
—
3----3 LCD
Tests LCD for possible problems by illuminating and extinguishing all LCD dots. While test continues, the LCDs are
forcibly switched in and out. Performance test consist of all dots lit (3 sec) and all dots unlit (3 sec) conducted twice
alternately and ends automatically (a total of sec). After the test, the LCDs restore normal illumination.
—
No.
Item
Description
Setup
Default
3---
Static
Runs a check for stability of sensor signal at zero flow.
For details, see .. Pipeline oscillation check (zero flow).
See ..
—
3---
Dynamic
Runs a check for stability of sensor signal while allowing the fluid.
For details, see .. Pipeline oscillation check (fluid flowing).
See ..
—
No.
Item
Description
Setup
Default
3--
Fix Analog
Produces a simulated output of analog output . It is beneficial for testing cable fault and performance of
associated equipment.
For details, see .7. Analog output.
See .7.
—
3--
Fix Analog
Produces a simulated output of analog output . It is beneficial for testing cable fault and performance of
associated equipment.
For details, see .7. Analog output.
See .7.
—
3--3
Fix Pulse
Produces a simulated output of pulse output . It is beneficial for testing cable fault and performance of
associated equipment.
For details, see .7. Pulse output.
See .7.
—
3--
Fix Pulse
Produces a simulated output of pulse output . It is beneficial for testing cable fault and performance of
associated equipment.
For details, see .7. Pulse output.
See .7.
—
3--
Status Output
Produces a simulated output of status output. It is beneficial for testing cable fault and performance of
associated equipment.
For details, see .7.3 Status output.
See .7.3
—
3--
Status Input
Monitors status input. It is beneficial for testing cable fault and performance of associated equipment.
For details, see .7. Status monitor input.
See .7.
—
No.
Item
Description
Setup
Default
3-3-
Auto Zero
Performs zeroing. When you perform zeroing, verify that the flow is at zero.
For details, see . Zeroing Function.