L-737-3-E
No.
Item
Description
Setup
Default
---
Assign
Changes analog output assignment. (See .3 Analog Output Function.)
If assignment is changed, reconfigure upper limit [No. ---], lower limit [No. ---3], low cutoff [No. --
-], and damping factor [No. ---] for optimum settings.
See . Mass Flow
---
URV
Changes analog output upper limit. A 0mA output is produced at the value so set.
A value smaller than lower limit [No. ---3] and a value larger than max. allowable range are not
acceptable.
See .3 0.000
---3
LRV
Changes analog output lower limit. A mA output is produced at the value so set.
A value larger than upper limit [No. ---] and a value larger than the max. allowable range are not
acceptable.
See .3 0.00000
---
Lowcut
Changes analog output low cutoff (%).
When low flow cutoff is greater than proportion of analog output where upper limit [No. ---] is set at
00% and lower limit [No. ---3] at 0%, analog output is clamped at mA.
See .3 0.0
---
Added Damp
Changes analog output damping factor (sec).
If analog output fluctuates excessively, select a larger value; if fast response is desired, select a smaller
value.
See .3 0.0
No.
Item
Description
Setup
Default
---
Assign
Changes analog output assignment. (See .3 Analog Output Function.)
If assignment is changed, reconfigure upper limit [No. ---], lower limit [No. ---3], low cutoff [No. --
-], and damping factor [No. ---] for optimum settings.
See . Mass Flow
---
URV
Changes analog output upper limit. A 0mA output is produced at the value so set.
A value smaller than lower limit [No. ---3] and a value larger than the max. allowable range are not
acceptable.
See .3 0.000
---3
LRV
Changes analog output lower limit. A mA output is produced at the value so set.
A value larger than upper limit [No. ---] and a value larger than the max. allowable range are not
acceptable.
See .3 0.00000
---
Lowcut
Changes analog output low cutoff (%).
When low flow cutoff is greater than proportion of analog output where upper limit [No. ---] is set at
00% and lower limit [No. ---3] at 0%, analog output is clamped at mA.
See .3 0.0
---
Added Damp
Changes analog output damping factor (sec).
If analog output fluctuates excessively, select a larger value; if fast response is desired, select a smaller
value.
See .3 0.0
No.
Item
Description
Setup
Default
--3-
Assign
Changes pulse output assignment. (For functions, see . Pulse Output Function.)
If assignment is changed, reconfigure frequency at full scale [No. --3-], full scale flowrate [No. --3-3],
and low cutoff [No. --3-] for optimum settings.
See . Mass Flow
--3-
Freq Factor
Changes the max. frequency of pulse output . An output is produced at the frequency (duty 0%) set at
full scale [No. --3-3]. Set up to match the measuring frequency range of instrument used for pulse output
measurement.
See .3 000.0
--3-3
Rate Factor
Changes pulse output full scale flowrate.
An output is produced at the max. frequency [No. --3-] when preset flowrate is reached.
See .3 0.000
--3-
Lowcut
Changes low cutoff (%) of pulse output .
When proportion of flowrate falls below low cutoff where full scale flowrate [No. --3-3] is set at 00%, the
pulse output is clamped at 0.
See .3 0.0
No.
Item
Description
Setup
Default
---
Assign
Changes pulse output assignment. (For functions, see . Pulse Output Function.)
If assignment is changed, reconfigure frequency at full scale [No. ---], full scale flowrate [No. ---3],
and low cutoff [No. ---] for optimum settings.
See . Mass Flow
---
Freq Factor
Changes the max. frequency of pulse output . An output is produced at the frequency (duty 0%) set at full
scale rate [No. ---3]. (If double pulse output is your option at pulse output assignment [No. ---],
this setting becomes invalid, however.)
See .3 000.0
---3
Rate Factor
Changes pulse output full scale flowrate. An output is produced at max. freq. [No. ---] when preset
rate is reached. (If double pulse output is your option at pulse output assignment [No. ---], this setting
becomes invalid, however.)
See .3 0.0
---
Lowcut
Changes pulse output low cutoff (%).
When proportion of flowrate falls below low cutoff where full scale flowrate [No. --3-3] is set at 00%, the
pulse output is clamped at 0 (zero).
See .3 0.0