| © Ottobock
Alignment and adjustment | Kenevo 3C60 | Prothetik | Global Academy | 12.12.2016
94
Alignment and adjustment
Waking up the knee joint : No connection with tube adapter
If there is no connection to the tube adapter
during wake up, Kenevo activates a safety
mode
•
Clear error:
−
Connect the tube adapter
−
Attach and remove charger (triggers a
reset/self-test)
→ wait for confirmation via feedback