Error message
Error no.
Content
Cause
Possible measure
M1 Brake Error
1505
Brake error on left
motor
M2 Brake Error
1506
Brake error on right
motor
e.g. defective
plug
connection,
cable
break
Defective brake
Check plug connec
tions and cable to
brake
If error persists: con
tact specialist dealer
M1 Motor Error
3B00
Left motor not con
nected
M2 Motor Error
3C00
Right motor not con
nected
e.g. defective
plug
connection,
cable
break
Check plug connec
tions and cable to
motor
Inhibit Active
1E01; 1E20; 1E21;
1E22; 1E23
Driving
function
blocked
due
to
external factors
Battery charger may
be connected
Disconnect
battery
charger
Brake Lamp Short
(None)
Short circuit in brake
light electrical circuit
e.g. defective
plug
connection,
cable
break
Defective brake light
Check plug connec
tions and cable to
brake light
If error persists: con
tact specialist dealer
Left Lamp Short
7205
Short circuit in elec
trical circuit of left
front/rear light
Right Lamp Short
7209
Short circuit in elec
trical circuit of right
front/rear light
e.g. defective
plug
connection,
cable
break
Defective
front/rear
light
Check plug connec
tions and cable to
front/rear light
If error persists: con
tact specialist dealer
L Ind Lamp Short
7206
Short circuit in elec
trical circuit of left
direction indicator
R Ind Lamp Short
720A
Short circuit in elec
trical circuit of right
direction indicator
e.g. defective
plug
connection,
cable
break
Defective
direction
indicator signal
Check plug connec
tions and cable to
direction indicator
If error persists: con
tact specialist dealer
L Ind Lamp Failed
7207
Error in electrical cir
cuit of left direction
indicator
R Ind Lamp Failed
7208
Error in electrical cir
cuit of right direction
indicator
e.g. defective
plug
connection,
cable
break
Defective bulb
Check plug connec
tions and cable to
direction indicator
If error persists: con
tact specialist dealer
Over-current
(None)
Amount of current in
an actuator circuit
too great
e.g. defective
plug
connection,
cable
break
Defective servomotor
(actuator) or control
ler
Check plug connec
tions and cable to
servomotor;
verify
that actuating drive
can move freely
If error persists: con
tact specialist dealer
Overtemp. (Acts)
(None)
Temperature on actu
ator circuits on con
troller too high
e.g. defective
plug
connections
Defective servomotor
(actuator)
Cool down phase
Check plug connec
tions and cable to
actuator
In case of recur
rence: contact spe
cialist dealer
Overtemp. (Lamps)
(None)
Temperature on the
light circuit on the
controller too high
e.g. defective
plug
connections
Defective lamp
Cool down phase
Check plug connec
tions and cable to
light
In case of recur
rence: contact spe
cialist dealer
107
Juvo
Maintenance and repair
Summary of Contents for Juvo
Page 1: ...Juvo Instructions for use user 3...
Page 2: ...2 Juvo...
Page 120: ...Juvo 120...
Page 121: ...121 Juvo...
Page 122: ...Juvo 122...