17
CAUTION
Remaining, operating and charging in areas outside the allowable temperature range
Injury due to faulty control or malfunction of the product.
►
Avoid remaining, operating and charging in areas outside the allowable temperature range
(see page 20).
CAUTION
Proximity to sources of strong magnetic or electrical interference (e.g. theft prevention
systems, metal detectors)
Injury due to unexpected behaviour of the product caused by interference with internal data
communication.
►
Avoid remaining in the vicinity of visible or concealed theft prevention systems at the
entrance/exit of stores, metal detectors/body scanners for persons (e.g. in airports) or other
sources of strong magnetic and electrical interference (e.g. high-voltage lines, transmitters,
transformer stations, computer tomographs, magnetic resonance tomographs, etc.).
►
When walking through theft prevention systems, body scanners or metal detectors, watch
for unexpected behaviour of the product.
CAUTION
Independent manipulation of the product
Injury due to malfunction and resulting unexpected prosthesis actions.
►
Manipulations to the product other than the tasks described in these instructions for use are
not permitted.
►
The product and any damaged components may only be opened and repaired by authorised,
qualified Ottobock personnel.
CAUTION
Charging the product with soiled or damaged contacts
Injury due to unexpected behaviour of the product caused by insufficient charging.
►
Make sure that the contacts are always clean and free of grease.
►
Clean the electrical contacts of the charging plug and charging receptacle regularly using
cotton swabs and a mild soap solution.
►
Take care to avoid damaging the contacts with pointed or sharp objects.
CAUTION
Distance to HF communication devices is too small (e.g. mobile phones, Bluetooth
devices, WiFi devices)
Injury due to unexpected behaviour of the product caused by interference with internal data
communication.
►
Therefore, keeping the following minimum distances to these HF communication devices is
recommended:
• Mobile phone GSM 850/GSM 900: 0.99 m
• Mobile phone GSM 1800/GSM 1900/UMTS: 0.7 m
• DECT cordless phones incl. base station: 0.35 m
• WiFi (routers, access points,…): 0.22 m
• Bluetooth devices (third-party products not approved by Ottobock): 0.22 m
Summary of Contents for 757B35 0
Page 2: ...1 2 3 4 5 6 7 8 2...
Page 3: ...9 10 3...
Page 126: ...126 757B35 0 757B35 1 757B35 3 3 3 1 3 2 132 3 3 3 4 Ottobock 4 4 1 4 2 1 2...
Page 127: ...127 4 3 Ottobock...
Page 128: ...128 5 Ottobock DermaClean 453H10 1 Ottobock 4 4 132...
Page 131: ...131 2 757B35 3 757B35 1 7 7 1 1 1 7 2 1 134 2 7 3 1 2 LED 7 4...
Page 135: ...135 1 2020 12 22 MyoEnergy Integral 757B35 2 2 1 2 1 1 1 2 3 1 2 1 2 3 2 1 2 3 3 1 2 3 4 1 2 5...
Page 136: ...136 757B35 0 757B35 1 757B35 3 3 3 1 3 2 143 3 3 3 4 Ottobock 4 4 1 4 2 1 2...
Page 137: ...137 4 3 Ottobock...
Page 138: ...138 5 Ottobock DermaClean 453H10 1 Ottobock 4 4...
Page 139: ...139 143 Ottobock Bluetooth WLAN...
Page 141: ...141 4 757B35 3 757B35 1 8 5 1 10 2 757B35 3 757B35 1 7 7 1 1 1 7 2 1 144 2 7 3 1 2 7 4...
Page 147: ...147 4 2 1 1 2 4 3...
Page 148: ...148 5 453H10 1 453 10 1 4 4...
Page 149: ...149 152 CT MRI WiFi...
Page 151: ...151 2 757B35 3 757B35 1 7 7 1 1 1 7 2 1 154 2 7 3 1 1 2 LED 7 4...
Page 154: ...154 11 2 11 2 1 LED LED 50 LED 11 2 2 1 2 LED 3 7in1 9E420 1 2 2...
Page 155: ...155...