– Adjusting/reading the 4 … 20 mA interface – lower limit
aOPA<value>!
aOPA<value<cr>
Adjust lower limit
aOPA!
aOPA<value<cr>
Read out lower limit
a
– Sensor address
<value>
–
pb.a
p
– Sign (+,–)
b
– Digits prior to the decimal point
b
– Digits after the decimal point
Number of spaces: 6 (5 decimal point)
Note the following settings for the units
m
→
pbb.aaa
Input in m (–30.000 … +30.000)
cm
→
pbb.aaa
Input in m (–30.000 … +30.000)
ft
→
pbb.aaa
Input in feet (–98.424 … +98.424)
Note: When switching between m and ft, rounding
errors of ±0.001 are possible.
– Adjusting/reading the 4 … 20 mA interface – upper limit
aOPB<value>!
aOPB<value<cr>
Adjust upper limit
aOPB!
aOPB<value<cr>
Read out upper limit
a
– Sensor address
<value>
–
pb.a
p
– Sign (+,–)
b
– Digits prior to the decimal point
b
– Digits after the decimal point
Number of spaces: 6 (5 decimal point)
Note the following settings for the units
m
→
pbb.aaa
Input in m (–30.000 …+30.000)
cm
→
pbb.aaa
Input in m (–30.000 … +30.000)
ft
→
pbb.aaa
Input in feet (–98.424 … +98.424)
Note: When switching between m and ft, rounding
errors of ±0.001 are possible.
– Adjusting/reading the 4 … 20 mA interface – mode
aOPC<value>!
aOPC<value<cr>
Adjust mode
aOPC!
aOPC<value<cr>
Read out mode
a
– Sensor address
<value>
–
M
ode
0
= 4 … 20 mA interface deactivated
(only the SDI-12 interface is active)
1
= 4 … 20mA and SDI-12 interface are active
2
= 4 mA output (constant value)
3
= 20 mA output (constant value)
4
= 12 mA output (constant value)
13