8285-H Jan 2020
Figure 8 – System Bypass
b. To allow contrast saving to resume, orient the DyeVert Power XT
Stopcock such that it is ON to both the
Valve Selection Stopcock
and
the catheter. (See Figure 7)
5) Contrast Accounting Estimation
a. At the end of the case, to estimate the total volume of contrast
that was delivered to the patient (i.e. Estimated Patient Contrast
Volume), subtract the Estimated Diverted Contrast Volume
(collected in the Contrast Collection Bag) from the power injector
Contrast Volume stated on the power injector display screen.
Estimated Patient Contrast Volume = [Contrast Volume injected
by power injector as shown in injector display – estimated
Diverted Contrast Volume contained in the Contrast Collection
Bag]
Note: contrast accounting estimation may be
inaccurate due to other fluids, such as blood or
saline being in the Contrast Collection Bag
.
Packaging Symbol Definitions
Expiration Date
YYYY-MM. Use
by last day of
month (MM).
Manufacturer
STERILE │EO
Sterilized by
ethylene oxide
Consult
electronic
Instructions for
Use
Keep Dry
REF
Model number
Single Use
European
Conformity
LOT
Lot Number
RxOnly
Prescription
Only
Do Not Use if
Package is
Damaged
EC │REP
European
Community
Authorized
Representative
Medical Device
Consult instructions
for use for
important
information
Identification of
Sterile Barrier
MR Unsafe
Power XT Module
Temperature
Limit
Osprey Medical Inc.
5600 Rowland Road, Suite 250
Minnetonka, MN 55343
USA
Customer Service Toll-Free:
1-855-860-7584 Fax: 1-855-883-4365
www.ospreymed.com
Osprey & DyeVert are trademarks of Osprey Medical Inc. ACIST CVi
®
and MEDRAD
®
Avanta™ are trademarks of their respective owner.
©Osprey Medical, Inc. 2020. All Rights Reserved.
DISCLAIMER OF WARRANTY AND LIMITATION OF REMEDY
THERE IS NO WARRANTY EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION
ANY IMPLIED WARRANTY OF MERCHANTABILITY OR OF FITNESS FOR A
PARTICULAR PURPOSE, ON THE OSPREY MEDICAL PRODUCT(S) DESCRIBED IN
THIS PUBLICATION. IN THE EVENT OF ANY DEFECT OR NONCONFORMITY OF OR TO
THIS PRODUCT(S), OSPREY MEDICAL’S LIABILITY SHALL NOT BE IN EXCESS OF THE
PURCHASE PRICE OF THE PRODUCT(S) TO BUYER. UNDER NO CIRCUMSTANCES
SHALL OSPREY MEDICAL BE LIABLE FOR ANY DIRECT, INDIRECT OR
CONSEQUENTIAL DAMAGES BASED UPON BREACH OF WARRANTY, BREACH OF
CONTRACT, NEGLIGENCE, STRICT TORT OR ANY OTHER THEORY ARISING OUT OF
THE PURCHASE, USE OR REUSE OF THIS PRODUCT(S). OSPREY MEDICAL NEITHER
ASSUMES NOR AUTHORIZES ANY PERSON TO ASSUME FOR IT ANY OTHER OR
ADDITIONAL LIABILITY OR RESPONSIBILITY IN CONNECTION WITH OSPREY
MEDICAL PRODUCT(S). Descriptions or specifications in Osprey Medical printed matter,
including this publication, are meant solely to generally describe the product at the time of
manufacture and do not constitute any express warranties.
2797
MedPass SAS
95 bis Bd Pereire, 75017 Paris, France
Australian Sponsor
Osprey Medical, Pty
Level 13, 41 Exhibition Street
Melbourne, Victoria 3000, Australia