Guidance and Manufacturer’s Declaration — ELECTROMAGNETIC EMISSIONS
LifePort Kidney Transporter is intended for use in the electromagnetic environment specified below. The customer or the
user of the LifePort should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic Environment:
Guidance
RF emissions
CISPR11
Group 1
LifePort Kidney Transporter uses
RF energy only for its internal
function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR11
Class A
The LifePort is suitable for use in all
establishments, other than domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Complies
Guidance and Manufacturer’s Declaration — ELECTROMAGNETIC IMMUNITY
LifePort Kidney Transporter is intended for use in the electromagnetic environment specified below. The customer or the
user of LifePort Kidney Transporter should assure that it is used in such an environment.
Immunity Test
IEC 60601 test level
Compliance level
Electromagnetic
Environment: Guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
Electrical fast transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±2 kV for power supply
lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV common mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70% UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
<5 % UT
(>95 % dip in UT)
for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70% UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of LifePort Kidney Transporter
requires continued operation
during mains power interruptions,
LifePort Kidney Transporter can
be lowered from the internal
battery.
Power frequency (50-60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
UT is the AC mains voltage prior to application of the test level.