LifePort Kidney Transporter Operator’s Manual 755-00002 Rev K
40
Closing LifePort Kidney Transporter
Having completed the visual inspection, close LifePort Kidney Transporter.
1. Replace the Inner Lid over the Perfusion Circuit.
2. Remove the Sterile Drape.
3.
According to hospital protocol, mark identification information on the Perfusion Circuit as indicated.
CAUTION:
Avoid marking identification on the Outer Lid. Identification on the Tubeframe is preferred.
4.
A person outside the aseptic field
, carefully place the Outer Lid on the Perfusion Circuit.
5.
A person outside the aseptic field
, close and latch the Outer Cover.
6. Recheck operating parameters on the Outer Display and record on the chart, per hospital procedure.
7. Verify that LifePort Kidney Transporter is in
INFUSE
mode.
8. Attach a handle tag (or similar approved label) to place identification information on the outside of
LifePort Kidney Transporter. Do not make permanent markings on LifePort Kidney Transporter unit itself.
NOTE:
LifePort Kidney Transporter is designed for unattended use, with no need for continuous
monitoring of the kidney.
Monitoring Options for a Kidney on LifePort Kidney Transporter
Once LifePort Kidney Transporter has been closed, there are several options for continuous monitoring of
the kidney until removal and transplantation into the recipient.
NOTE:
You can also connect LifePort Kidney Transporter USB-A port to a computer and download
operating parameters. For more information, refer to
Capturing and Downloading Data
on page 48.
Step 1
Step 5
Step 6