LifePort Kidney Transporter Operator’s Manual 755-00002 Rev K
36
7.
Using aseptic technique
, place the Kidney Cradle in the Perfusion Circuit.
Priming the Infuse Line
When the Kidney Cradle containing the kidney has been placed in the Perfusion Circuit, use the
following procedure to prime the Infuse Line, removing bubbles from the line and renal artery. This
procedure is performed using aseptic technique:
NOTE:
For better clarity, the following illustrations of the Perfusion Circuit and Kidney Cradle with kidney
will be shown outside LifePort Kidney Transporter.
1.
Connect the Infuse Line as shown and tighten the Luer Lock fitting.
Step 7
Connect the
Infuse Line
With the SealRing cannula:
Connect the
Infuse Line
With the Straight cannula:
With the Coupler: