755-00002 Rev K
LifePort Kidney Transporter Operator’s Manual
33
4. Lower the upper portion of the cannula carefully bringing the conical seal into the interior of the
vessel. Viewing through the magnified sight-glass, verify that the vessel orifice is centered,
circular in shape and not occluded. Wrap each strap, straight and securely, around both
cannula halves and fix the straps to their posts.
5.
Gravity-flow small amounts of flush solution into the cannula, then check for and repair any
leaks that occurred in the surgical process or with the cannula.
6. Place the kidney in the Kidney Cradle and snap the cannula into the Cannula Mount.
7. Adjust the height and rotation of the Cannula Mount to comfortably position the vessel.
8. Visually inspect the vessel, ensuring there are no twists or occlusions.
9. Drape the Mesh Organ Restraint over the kidney and secure the organ to the Kidney Cradle.
Step 8
Step 9
Step 5
Step 6