48
-
Appendix C
– Smart Averaging
The average function is generally used to smoothen the output signal. With the adjustable parameter time
this function can be optimal adjusted to the respective application. One disadvantage of the average function
is that fast temperature peaks which are caused by dynamic events are subjected to the same averaging
time. Therefore those peaks can only be seen with a delay on the signal output.
The function
Smart Averaging
eliminates this disadvantage by passing those fast events without averaging
directly through to the signal output.
Signal graph with Smart Averaging function
Signal graph without Smart Averaging function