General Information
9
-
1.6 Factory Default Settings
The unit has the following presetting at time of delivery:
Signal output object temperature
4-20 mA
Emissivity (switches)
0,970 [LT/ hs LT]
1,000 [2ML/ 2MH/ G5HF]
Emissivity (via software)
1,000
Transmissivity
1,000
Average time (AVG)
0,2 s [LT, hs LT]
0,1 s [2ML, 2MH, G5HF]
Smart Averaging
inactive [LT]
active [hs LT, 2ML, 2MH, G5HF]
Peak hold
inactive
Valley hold
inactive
LT
hs LT
2ML
2MH
G5HF
Lower limit temperature range [°C]
0
-20
250
385
200
Upper limit temperature range [°C]
500
150
800
1600
1650
Lower limit signal output
4 mA
Upper limit signal output
20 mA
Temperature unit
°C
Ambient temperature compensation
internal head temperature probe
Laser
active
Summary of Contents for CSlaser 2MH
Page 1: ...Operator s Manual optris CSlaser LT hs LT 2M G5 Infrared thermometer...
Page 23: ...Technical Data 23 G5HF Optics CF4 D S focus distance 45 1 10mm 450mm D S far field 15 1...
Page 49: ...Appendix D Declaration of Conformity 49 Appendix D Declaration of Conformity...
Page 50: ...optris CSlaser MA E2018 12 A...