Electrical Installation
31
-
Connector version
This version has a connector plug integrated in the sensor backplane.
Therefore an opening of the sensor for cable assembling is not necessary.
Please use the original ready-made, fitting connection cables which are
optionally available. If you want to use own cables please note the pin
assignment of the connector (see next page).
Power supply
Please use a power supply unit with an output voltage of
5
–30 VDC
which can supply
100 mA.