Accessories
27
-
4.3 Water Cooled Housing
Water cooled housing [ACCTLW]
Hose connection: 6x8 mm
Thread (fitting): G 1/8 inch
The CSlaser can be used at ambient temperatures up to 85 °C without cooling. For applications, where the
ambient temperature can reach higher values, the usage of the optional water cooled housing is
recommended (operating temperature up to 175 °C). The sensor should be equipped with the optional high
temperature cable (operating temperature up to 180 °C).
To avoid condensation on the optics an air purge
collar is recommended.
Water flow rate: approx. 2 l/ min
(Cooling water temperature should not exceed
30 °C)