36
-
6 Emissivity Setting
After opening of the sensor backplane
[►5.1 Cable Connections]
both of the emissivity switches are
accessible.
For an emissivity setting of
1,00
please turn both switches to
0
. Values below
0,10
are not adjustable. For all
other switch positions the following applies:
0, S1 S2
.
Therefore the adjustment range is
0,10...1,09
.
Example:
ε = 0,84
S1=8
S2=4
If you use the software (optional) please consider that the emissivity
switches can be activated/ deactivated in the software menu
Device/
Device setup
. At time of delivery the switches are active.
The emissivity set in the software interacts as a factor to the emissivity
set on the unit. Thus the adjustment range increases to
0,100...1,199
.
Example:
ε
Software
= 0,952 x
ε
Sensor
= 0,82 (S1=8/ S2=2)
Therefore the effective emissivity is: 0,781.