5. Troubleshooting 

If there is any trouble occurs in operation, please referring to the following sheet listed 

some common troubleshooting resolve them or keep in touch with us or local 

agency

Trouble 

Causation 

Remedy 

Switch on but bulb 

dark 

Plug is unreliable 

Check joint again 

Bulb is broken 

Replace bulb 

Fuse is broken 

Replace fuse 

Bulb is flickering 

or brightness is 

unsteady. 

Bulb is unstable 

Insert again 

Bulb is broken 

Replace bulb 

Brightness of view 

field isn’t enough 

or is uneven. 

Bulb specification doesn’t   
meet the requirement.

 

Replace bulb 

Bulb brightness is low. 

Rotate potentiometer to adjusting brightness.

 

Objective isn’t in correct optical path. 

Rotate the objective in correct position. 

The size of iris aperture is too 
Small.

 

Adjust the size of iris aperture. 

Lens (objective, eyepiece, 
condenser, light collector) has dust

 

Wipe Clean . 

Position of Condenser is too low   

Adjust condenser properly. 

Image    isn’t   

clear(contrast    or 

definition isn’t   

enough)

 

Cover glass of specimen 
doesn’t meet the requirement 

Use required thickness cover glass (0.17mm). 

Cover  glass  of  specimen  isn’t  in  up 

direction 

Place specimen correctly. 

There is bubble in immersion oil 

Clear the bubble away 

Size of iris aperture isn’t   
proper 

Adjust the size of iris aperture. 

Position of condenser is too low 

Readjust again 

One side of image 

is dark or image is 

moving as 

focusing. 

Objective isn’t in correct 
optical path.

 

Make the objective in correct position. 

Specimen isn’t placed correctly. 

Place  specimen  levelly  on  stage  and  clip  it 

with clamp. 

Objective touching 

specimen as 

changing low 

power to high 

power

 

Cover glass of specimen 
doesn’t meet the requirement. 

Place specimen correctly. 

Cover glass is too thick. 

Use required thickness cover glass (0.17mm). 

It is easy for eyes 

to be tired during 

observing. 

Illumination brightness isn’t enough. 

Adjust brightness 

Diopter isn’t adjusted correctly. 

Adjust diopter 

Contrast isn’t enough 

Adjust aperture diaphragm. 

Summary of Contents for A15.2604

Page 1: ...A15 2604 Polarizing Microscope To ensure the safety and obtain satisfactory performance please study this instruction manual thoroughly before start to use the instrument...

Page 2: ...vent a fire 4 Notes on replacing the bulb The instrument used with 12V 20W imported halogen lamp of G4 replacement should be used the same specifications of the bulb or it may cause equipment damage B...

Page 3: ...component 7 3 2 2 Place sample 7 3 2 3 Condenser 7 3 2 4 Focusing 7 3 2 5 Polarizing observation 7 3 2 6 Use of tension adjusting handle 9 3 2 7 Notice after operation 9 4 Maintenance 9 4 1 Clean 9 4...

Page 4: ...pe Size A 01 Monocular tube 02 Slip 4 Slip 03 Nosepiece 04 Stage 05 Condenser 06 Filter removable 07 Collector 08 Base 09 Condenser lifting handle 10 Left Fine focusing knobs 11 Left Coarse focusing k...

Page 5: ...w 17 Analyzer change over handle 18 Vernier 19 Stage locking screw 20 Aperture diaphragm handle 21 Polarizer unit 22 Potentiometer knob 23 Right Coarse focusing knobs 24 Right fine focusing knobs 25 T...

Page 6: ...0 400 200 600 10 13 Optional 40 100 400 200 600 2 2 Objective with Eyepiece 10 18 Non stress Achromatic Objective Numerical Aperture N A Objective Field of View Resolution 4 0 10 4 5mm 3 36 m 10 0 25...

Page 7: ...Measures 2 3 12 Electric components Input voltage AC100 240V 47 63Hz Output voltage DC3V 12V Halogen lamp 12V 20W LED lamp is optional Potentiometer with switch rotate knob Fuse 5A 5 20 3 Installatio...

Page 8: ...Adjusting the distance between condenser and sample by rotating condenser lifting handle can change the light s degree of uniformity for best brightness b Adjustment of Aperture Diaphragm Adjust the s...

Page 9: ...s of light absorption color diversity optical properties related with refractive index processus exasperate edge baker red dispersion etc e Angular deflection of analyzer can enhance contrast when in...

Page 10: ...after operation a Please cut off power rotate potentiometer knob when not in use unplug the power socket If use immersion oil should wipe off the lens and slices put on dust cover b If not in use for...

Page 11: ...life and insert fully the contact pins into the bulb socket 4 3 7 Press back the lamp housing cover and tighten the knurled thumb screw After working for above 8 hours continuously should cut off the...

Page 12: ...er is too low Adjust condenser properly Image isn t clear contrast or definition isn t enough Cover glass of specimen doesn t meet the requirement Use required thickness cover glass 0 17mm Cover glass...

Page 13: ...12 B Pol Polarizing Microscope...

Reviews: