1. Purposes 

This  transmitting  polarizing  microscope  is  designed  on  the  basis  of  biological 

microscope ,use the polarization of light to make microscopic observation for double 
refraction  material.  It  is  widely  used  in  geology,  chemical  industry  ,metallurgy, 
semiconductor,  petroleum,  spinning  field  etc,  also  play  important  role  in  biological, 
phytology,

 

pharmacology, pathology field etc. 

  

2. Performance Parameter 

2.1

Total Magnification

   

          Objective 

Eyepiece 

4× 

10× 

40× 

20×

optional

 

60×

optional

 

10×/φ18(Reticule) 

40× 

100× 

400× 

200× 

600× 

10×/φ13(Optional) 

40× 

100× 

400× 

200× 

600× 

 

2.2

.

 

Objective

with Eyepiece 10×/18

 

Non-stress   

Achromatic 

Objective 

Numerical 

Aperture 

  N.A. 

Objective 

Field of View 

Resolution 

4× 

0.10 

4.5mm 

3.36μm 

10× 

0.25 

1.8mm 

1.34μm 

40×(s) 

0.65 

0.45mm 

0.52μm 

20×

optional

 

0.40 

0.9mm 

0.84μm 

60×(s)

optional

 

0.80 

0.3mm 

0.42μm 

 
2.3.    Other Parameter 

2.3.1. 

Mechanical Tube Length

      160mm 

2.3.2. 

Conjugate distance

        195mm 

2.3.3. 

Eyepiece viewing tube

  Monocular head

Inclined 30°. 

2.3.4. 

Midbody

 

  Analyzer can rotate 90°

with scale, can move out of the optical path for single 

polarizing observation.   

  Bertrand lens can be move out of the optical path 

Standard equipped λ and λ/4 

quartz wedge is optional. 

2.3.5.

Nosepiece

  Quadplex hole inner location. 

Summary of Contents for A15.2604

Page 1: ...A15 2604 Polarizing Microscope To ensure the safety and obtain satisfactory performance please study this instruction manual thoroughly before start to use the instrument...

Page 2: ...vent a fire 4 Notes on replacing the bulb The instrument used with 12V 20W imported halogen lamp of G4 replacement should be used the same specifications of the bulb or it may cause equipment damage B...

Page 3: ...component 7 3 2 2 Place sample 7 3 2 3 Condenser 7 3 2 4 Focusing 7 3 2 5 Polarizing observation 7 3 2 6 Use of tension adjusting handle 9 3 2 7 Notice after operation 9 4 Maintenance 9 4 1 Clean 9 4...

Page 4: ...pe Size A 01 Monocular tube 02 Slip 4 Slip 03 Nosepiece 04 Stage 05 Condenser 06 Filter removable 07 Collector 08 Base 09 Condenser lifting handle 10 Left Fine focusing knobs 11 Left Coarse focusing k...

Page 5: ...w 17 Analyzer change over handle 18 Vernier 19 Stage locking screw 20 Aperture diaphragm handle 21 Polarizer unit 22 Potentiometer knob 23 Right Coarse focusing knobs 24 Right fine focusing knobs 25 T...

Page 6: ...0 400 200 600 10 13 Optional 40 100 400 200 600 2 2 Objective with Eyepiece 10 18 Non stress Achromatic Objective Numerical Aperture N A Objective Field of View Resolution 4 0 10 4 5mm 3 36 m 10 0 25...

Page 7: ...Measures 2 3 12 Electric components Input voltage AC100 240V 47 63Hz Output voltage DC3V 12V Halogen lamp 12V 20W LED lamp is optional Potentiometer with switch rotate knob Fuse 5A 5 20 3 Installatio...

Page 8: ...Adjusting the distance between condenser and sample by rotating condenser lifting handle can change the light s degree of uniformity for best brightness b Adjustment of Aperture Diaphragm Adjust the s...

Page 9: ...s of light absorption color diversity optical properties related with refractive index processus exasperate edge baker red dispersion etc e Angular deflection of analyzer can enhance contrast when in...

Page 10: ...after operation a Please cut off power rotate potentiometer knob when not in use unplug the power socket If use immersion oil should wipe off the lens and slices put on dust cover b If not in use for...

Page 11: ...life and insert fully the contact pins into the bulb socket 4 3 7 Press back the lamp housing cover and tighten the knurled thumb screw After working for above 8 hours continuously should cut off the...

Page 12: ...er is too low Adjust condenser properly Image isn t clear contrast or definition isn t enough Cover glass of specimen doesn t meet the requirement Use required thickness cover glass 0 17mm Cover glass...

Page 13: ...12 B Pol Polarizing Microscope...

Reviews: