Conoscopic  observation  usually  with  high  power  objective  and  can  clearly  see  the  amplified 

interference pattern, confirm its axis and slice direction, determine mineral’s light symbol and 

axis angle (biaxial crystal

▲Don’t reverse the direction forcefully for left and right coarse & fine focusing knobs at 
the same time ,otherwise will damage focusing unit. 
▲Transform    eyepiece with different power   

don’t direct wrench    objective to rotate 

nosepiece. should rotate the nosepiece by holding the insection of nosepiece cover to make 
objective  in  right  position    for  optical  path

otherwise  will  cause    effect  for  optical 

performance. 
3.2.6 Use of tension adjusting handle   

Stage will slide down automatically after in long term usage.   
Tightness adjusting ring can adjust coarse focusing unit in case of this conditions and make 
adjusting operation comfortable.                                                                                       
Clockwise rotation is loose, in contrast ,is locked. 

3.2.7 Notice after operation. 

a. Please cut off power (rotate potentiometer knob )when not in use, unplug the power socket. 

If use immersion oil ,should wipe off the lens and slices , put on dust cover. 

b. If not in use for long time, suggest to place eyepiece 

objective in the dry containers (such 

as moisture-proof cylinder), and place desiccant. Put on dust cover for main body.   

 

4. Maintenance 

4.1 Clean 

4.1.1 Don’t touch the lens with hand, Dust on lens should be cleaned by soft brush or absorbent 

cotton or cleaned by absorbent cotton, lens paper with the mixture of alcohol and ether 

(proportion 1:4). 

4.1.2 Alcohol and ether all are burnt easily, please take them away from fire. Be careful for turn 

on and off power. 

4.1.3 Don’t clean painted metal and galvanizing metal with organic solvent such as alcohol, 

ether or the mixture of the both. Silicon cloth or soft cleaning preparation is suggested to clean 

it.   

4.1.4 Plastic should be cleaned by soft cloth with clear water. 

4.2 Enviroment for use   

4.2.1 Microscope should be used and placed in a cool, dry, non-dust, non-shake and 
non-corrosive gases environment. 
4.2.2 Normal working condition

 

Summary of Contents for A15.2604

Page 1: ...A15 2604 Polarizing Microscope To ensure the safety and obtain satisfactory performance please study this instruction manual thoroughly before start to use the instrument...

Page 2: ...vent a fire 4 Notes on replacing the bulb The instrument used with 12V 20W imported halogen lamp of G4 replacement should be used the same specifications of the bulb or it may cause equipment damage B...

Page 3: ...component 7 3 2 2 Place sample 7 3 2 3 Condenser 7 3 2 4 Focusing 7 3 2 5 Polarizing observation 7 3 2 6 Use of tension adjusting handle 9 3 2 7 Notice after operation 9 4 Maintenance 9 4 1 Clean 9 4...

Page 4: ...pe Size A 01 Monocular tube 02 Slip 4 Slip 03 Nosepiece 04 Stage 05 Condenser 06 Filter removable 07 Collector 08 Base 09 Condenser lifting handle 10 Left Fine focusing knobs 11 Left Coarse focusing k...

Page 5: ...w 17 Analyzer change over handle 18 Vernier 19 Stage locking screw 20 Aperture diaphragm handle 21 Polarizer unit 22 Potentiometer knob 23 Right Coarse focusing knobs 24 Right fine focusing knobs 25 T...

Page 6: ...0 400 200 600 10 13 Optional 40 100 400 200 600 2 2 Objective with Eyepiece 10 18 Non stress Achromatic Objective Numerical Aperture N A Objective Field of View Resolution 4 0 10 4 5mm 3 36 m 10 0 25...

Page 7: ...Measures 2 3 12 Electric components Input voltage AC100 240V 47 63Hz Output voltage DC3V 12V Halogen lamp 12V 20W LED lamp is optional Potentiometer with switch rotate knob Fuse 5A 5 20 3 Installatio...

Page 8: ...Adjusting the distance between condenser and sample by rotating condenser lifting handle can change the light s degree of uniformity for best brightness b Adjustment of Aperture Diaphragm Adjust the s...

Page 9: ...s of light absorption color diversity optical properties related with refractive index processus exasperate edge baker red dispersion etc e Angular deflection of analyzer can enhance contrast when in...

Page 10: ...after operation a Please cut off power rotate potentiometer knob when not in use unplug the power socket If use immersion oil should wipe off the lens and slices put on dust cover b If not in use for...

Page 11: ...life and insert fully the contact pins into the bulb socket 4 3 7 Press back the lamp housing cover and tighten the knurled thumb screw After working for above 8 hours continuously should cut off the...

Page 12: ...er is too low Adjust condenser properly Image isn t clear contrast or definition isn t enough Cover glass of specimen doesn t meet the requirement Use required thickness cover glass 0 17mm Cover glass...