background image

 

 

- 3 -

 

Fig

.1 B-POL Polarizing Microscope Size -A 

 

01—Monocular tube                                  02—λ Slip , λ/4 Slip 
03—Nosepiece                                          04—Stage 
05—Condenser                                          06—Filter

removable

 

07—Collector                                            08—Base 
09—Condenser lifting handle                  10—Left Fine focusing knobs 
11—Left Coarse focusing knobs              12—Clamps 
13—Body                                                    14—Midbody 

 

Summary of Contents for A15.2604

Page 1: ...A15 2604 Polarizing Microscope To ensure the safety and obtain satisfactory performance please study this instruction manual thoroughly before start to use the instrument...

Page 2: ...vent a fire 4 Notes on replacing the bulb The instrument used with 12V 20W imported halogen lamp of G4 replacement should be used the same specifications of the bulb or it may cause equipment damage B...

Page 3: ...component 7 3 2 2 Place sample 7 3 2 3 Condenser 7 3 2 4 Focusing 7 3 2 5 Polarizing observation 7 3 2 6 Use of tension adjusting handle 9 3 2 7 Notice after operation 9 4 Maintenance 9 4 1 Clean 9 4...

Page 4: ...pe Size A 01 Monocular tube 02 Slip 4 Slip 03 Nosepiece 04 Stage 05 Condenser 06 Filter removable 07 Collector 08 Base 09 Condenser lifting handle 10 Left Fine focusing knobs 11 Left Coarse focusing k...

Page 5: ...w 17 Analyzer change over handle 18 Vernier 19 Stage locking screw 20 Aperture diaphragm handle 21 Polarizer unit 22 Potentiometer knob 23 Right Coarse focusing knobs 24 Right fine focusing knobs 25 T...

Page 6: ...0 400 200 600 10 13 Optional 40 100 400 200 600 2 2 Objective with Eyepiece 10 18 Non stress Achromatic Objective Numerical Aperture N A Objective Field of View Resolution 4 0 10 4 5mm 3 36 m 10 0 25...

Page 7: ...Measures 2 3 12 Electric components Input voltage AC100 240V 47 63Hz Output voltage DC3V 12V Halogen lamp 12V 20W LED lamp is optional Potentiometer with switch rotate knob Fuse 5A 5 20 3 Installatio...

Page 8: ...Adjusting the distance between condenser and sample by rotating condenser lifting handle can change the light s degree of uniformity for best brightness b Adjustment of Aperture Diaphragm Adjust the s...

Page 9: ...s of light absorption color diversity optical properties related with refractive index processus exasperate edge baker red dispersion etc e Angular deflection of analyzer can enhance contrast when in...

Page 10: ...after operation a Please cut off power rotate potentiometer knob when not in use unplug the power socket If use immersion oil should wipe off the lens and slices put on dust cover b If not in use for...

Page 11: ...life and insert fully the contact pins into the bulb socket 4 3 7 Press back the lamp housing cover and tighten the knurled thumb screw After working for above 8 hours continuously should cut off the...

Page 12: ...er is too low Adjust condenser properly Image isn t clear contrast or definition isn t enough Cover glass of specimen doesn t meet the requirement Use required thickness cover glass 0 17mm Cover glass...

Page 13: ...12 B Pol Polarizing Microscope...

Reviews: