Adjustment
22
Adjust and display the measurement values according to usage. In addition, adjust
and display the measurement values according to the environment where used.
To adjust, press
.
Caution
・
Pressing
automatically retains the measurement value.
・
If offset, span, or 2-point adjustment has already been
performed, be sure to initialize the adjustment before
performing the 2-point adjustment.
・
When “Display item (unit selection)” P.13 is changed, the values
of the offset adjustment, span adjustment, and 2-point adjustment
that have set the display item are reset to the factory default values.
9
Adjustment
Set the offset coefficient to add to a
measurement value.
Display the measurement value with
offset-adjusted to any display value.
For 4-20 mA output (“6 4-20 mA Setting” P.15)
and alarm output (“Alarm output setting” P.17),
the adjusted values are applied.
Offset adjustment
Measurement item [Display]
Setting range
Factory default setting
Turbidity [Turb.]
-300.0 to 300.0 0.0
Press
.
The measurement item selection display is displayed.
Select [Turb. TC-3000-D] using
, and press
.
1
2
$GMXVW
0HDVXUHPHQW,WHP6HOHFW
㸯
S+ࠉ6(S+'
㸰
7XUE7&'
㸱
'2'26
㸱
7HPS'26
Measurement item
Sensor model
Sensor channel
Display value
after offset
adjustment
Measurement
value
Use
to switch pages.
The adjustment menu is displayed.
TC-3000SP-D is also displayed as
[TC-3000-D].