Relay Setting
17
7
Relay Setting
Set the relay output setting from SC-U1.
The factory default setting is OFF for relay output.
To set the relay setting, press
.
Set the alarm to be output when the measurement value exceeds a certain
value.
The following settings are available.
Alarm output setting
Setting item
Setting range
Turbidity [Turb.]
Trigger
High Alarm/Low Alarm
Active Level
0.0 to 3000
Hysteresis
High Alarm: 0.0 to Active level
Low Alarm: 0.0 to (3000 - Active level)
ON Delay
OFF
1 to 120 minutes
OFF Delay
OFF
1 to 120 minutes
The setting unit for the active level is based on the display resolution of the item
set for the display item. (For display resolution, refer to “14 Specifications” P.34)
For operation, refer to the "Universal Transmitter SC-U1 Operation
Manual", "5.4 Relay Setting" and "
◆
Alarm output setting".
Caution
Pressing
automatically retains the measurement value.
Caution
When “Display item (unit selection)” P.13 is changed, the
output item of the channel of relay setting that has set the display
item becomes OFF.
Set again if necessary.