Power
12-28VDC
4-20mA Pulse
/Alarm
!
Earth Ground
+12 to 28VDC
12 to 28VDC Return
CAUTION!
• Supply connection wiring must be rated for at least 90°C.
Power Input Requirements: 12 to 28VDC Supply
External DC power supply must provide 12 to 28VDC (10 to 30VDC full input power
range) at 6 Watts minimum.
(With 12VDC power, the F-5500 can use up to 500mA. With 24VDC power, the F-5500
can use up to 250mA.)
A 20 Watt or greater power supply is recommended to ensure it can provide enough
current under all temperature, ventilation and power on conditions.
The enclosure must be properly grounded with a quality earth ground. Sixteen (16)
gauge, stranded wire, is recommended for power and earth ground.
Fig. 3.1: Connections for 12 to 28VDC Supply
11451 Belcher Road South, Largo, FL 33773 • USA • Tel +1 (727) 447-6140 • Fax +1 (727) 442-5699 • [email protected]
F-5500 Thermal Mass Flow Meter Manual 06/17 - 2030 / 107023
Page 30
Model F-5500
WIRING
Wiring: Input Power
Power Input
Wiring