(2)-6 Statistical Calculation Settings
The ZR-RX70 can perform two statistical calculations.
This section describes specifying the statistical calculation settings.
Selection
Description
Off
Calculation is not performed.
Average
Displays the simple average value of data being captured.
Max
Displays the maximum value of data being captured.
Min
Displays the minimum value of data being captured.
Peak
Displays the peak value of data being captured.
RMS
Displays the effective value of data being captured.
The calculation formula is as shown below.
D: data n: number of data
• The calculation results are displayed in the D Statistical Calculation Display screen. Switch to the Full
Ch Statistical Calculation Display using the and keys.
For details, see “(6) Direction keys” p. 47
• Calculation will start upon power ON. Pressing the [START] key to begin measurement will clear the calculation.
• The sampling speed for calculation corresponds to the speed set in the DATA menu. However, when the sampling
speed is higher than 5ms, it is set to 5ms.
Note
Summary of Contents for ZR-RX70
Page 1: ...User s Manual Portable Multi Logger ZR RX70 Cat No Z283 E1 02...
Page 12: ...10 ZR RX70 User s Manual MEMO...
Page 16: ...14 ZR RX70 User s Manual MEMO...
Page 108: ...WEB Server Function 106 ZR RX70 User s Manual MEMO...
Page 118: ...External Dimensions 116 ZR RX70 User s Manual MEMO...
Page 119: ...5 APPENDIX APPENDIX Error Messages and Countermeasures 118 Index 121 Revision History 124...
Page 122: ...Error Messages and Countermeasures 120 ZR RX70 User s Manual MEMO...
Page 125: ...Index ZR RX70 User s Manual 123 5 APPENDIX MEMO...