Description
HiTemp150-TSK
HiTemp150-PT-TSK
Temperature Sensor
100
Ω
platinum RTD
Temperature Range
–40 °C to +250 °C
Probe Temperature Range
-200 ºC to +260 ºC
Temperature Resolution
0.05 °C
Temperature Accuracy
+0.5 °C
Memory
32,767
Reading Rate
1 reading every second up to 1 reading every 12 hours
Required Interface Package
IFC110 or IFC200
Baud Rate
2,400
Typical Battery Life
1 year
Operating Environment
–200 °C to +250 °C (Time Limited), 0 %RH to 100 %RH
Material
Data Logger: 316 stainless steel
Thermal Shield: Teflon
Dimensions: Body
Body: 1.1 in x 1.75 in dia. (28 mm x 45 mm dia.)
Dimensions: Shield
Shield: 2.1 in x 2.6 in dia. (53.34 mm x 66 mm dia.)
Dimensions: Probe
Probe: 2.0 in x 3/16 in dia.
(50 mm x 5 mm dia.)
Tip: 0.125 in OD x 1.5 in L
Flexible Portion: 24 in x 0.062 in OD dia.
Weight
17 oz (481.9 g)
10.40 oz (294.84 g)
Approvals
CE
HiTemp150-TSK and
HiTemp150-PT-TSK
Red 485
Dark Blue Pantone 289
Light Blue 58% of Pantone 289
To view the full MadgeTech product line,
visit our website at
www.madgetech.com.
Red 485
Dark Blue Pantone 289
Light Blue 58% of Pantone 289
Battery Warning
WARNING: FIRE, EXPLOSION, AND SEVERE BURN HAZARD. DO NOT SHORT CIRCUIT, CHARGE, FORCE
OVER DISCHARGE, DISASSEMBLE, CRUSH, PENETRATE OR INCINERATE. BATTERY MAY LEAK OR
EXPLODE IF HEATED ABOVE 150 °C (302 °F).
Specifications subject to change.
See MadgeTech’s terms and conditions at www.madgetech.com
MadgeTech, Inc.
6 Warner Road
l
Warner, NH 03278
Phone 603.456.2011
l
Fax 603.456.2012
www.madgetech.com
l
DOC-1204035-00 REV 7 2014.09.30
HiTemp150-TSK
Extended Range Temperature Data Logger with Thermal Shield
Product Information Card
Ambient Temperature
Time in Air to Max Internal Temp
(150 °C/302 °F)
Time in Liquid to Max Internal Temp
(150 °C/302 °F)
-200 °C (-328 °F)
18 minutes
N/A
-180 °C (-292 °F)
19 minutes
N/A
-160 °C (-256 °F)
21 minutes
N/A
-140 °C (-220 °F)
24 minutes
N/A
-120 °C (-184 °F)
27 minutes
N/A
-100 °C (-148 °F)
32 minutes
N/A
-80 °C (-112 °F)
40 minutes
N/A
-60 °C (-76 °F)
55 minutes
25 minutes
-40 °C to +150 °C (-40 °F to +302 °F)
Indefinitely
Indefinitely
160 °C (320 °F)
75 minutes
34 minutes
170 °C (338 °F)
63 minutes
29 minutes
180 °C (356 °F)
55 minutes
26 minutes
190 °C (374 °F)
50 minutes
23 minutes
200 °C (392 °F)
45 minutes
21 minutes
210 °C (410 °F)
42 minutes
19 minutes
220 °C (428 °F)
39 minutes
18 minutes
230 °C (446 °F)
36 minutes
17 minutes
240 °C (464 °F)
34 minutes
16 minutes
250 °C (482 °F)
32 minutes
15 minutes
HiTemp150-PT-TSK
Extended Range Temperature Data Logger with a 24”
Stainless Steel Flexible Probe and Thermal Shield