![Omron ZR-RX70 User Manual Download Page 114](http://html1.mh-extra.com/html/omron/zr-rx70/zr-rx70_user-manual_742803114.webp)
Standard Specifications
112
ZR-RX70 User’s Manual
*1
A unit in
µ
s and External cannot be selected if the save destination is the internal flush memory or USB memory.
*2
Neither the internal flush memory nor USB memory can be selected if a unit in µs is selected.
*3
This function is available only if data is captured to the internal RAM.
*4
Realtime and between cursors specified (during data replay)
Other
functions
Search functions
Function:
Search the captured data for the required number of points
Search type: Search of channels by levels
Search by logic combinations
Search by alarm generation
Annotation input
function
Function: A comment can be input for each channel.
Supported characters: Alphanumeric and kana characters
Number of characters: 31 (Up to 8 characters are displayed)
Trigger
Functions
Timer mode
Off, Date and Time, Every Day Cycle, Every Hour Cycle
Repeat trigger
Off, On
Trigger types
Start: Data capture starts when a trigger is generated.
Stop: Data capture stops when a trigger is generated.
Trigger settings
Start: Off, Level, External, Time
Stop: Off, Level, External, Time
Level
judgment modes
Analog:
↑
H,
↓
L, Window IN, Window Out (Tolerance ±1%)
Logic:
↑
H,
↓
L
Pulse:
↑
H,
↓
L, Window IN, Window Out
Trigger accuracy
± 1% of the range
Channel
combinations
Level OR, Level AND, Edge OR, Edge AND
Pulse input
setting
Revolutions mode
Function: Counts the number of pulses per second and multiplies it by 60 to
display an rpm value.
Span:
50 100 200 500 1k 2k 5k 10k 20k 50k 100k 200k 500k
1M 2M 5M 10M 20M RPM/F.S.
Counts mode
Function: Displays a count of the number of pulses for each sampling interval
from the start of measurement.
Span:
50 100 200 500 1k 2k 5k 10k 20k 50k 100k 200k 500k
1M 2M 5M 10M 20M C/F.S.
Inst.mode
Function: Counts the number of pulses for each sampling interval.
Resets the count value after each sampling interval.
Span:
50 100 200 500 1k 2k 5k 10k 20k 50k 100k 200k 500k
1M 2M 5M 10M 20M C/F.S.
Maximum number of
pulse inputs
Maximum input frequency: 50 kHz
Maximum number of counts: 15 MC (24-bit counter)
External sampling input
Maximum input frequency : 100kHz
Time error :
10µs or less
Item
ZR-RX70A
Summary of Contents for ZR-RX70
Page 1: ...User s Manual Portable Multi Logger ZR RX70 Cat No Z283 E1 02...
Page 12: ...10 ZR RX70 User s Manual MEMO...
Page 16: ...14 ZR RX70 User s Manual MEMO...
Page 108: ...WEB Server Function 106 ZR RX70 User s Manual MEMO...
Page 118: ...External Dimensions 116 ZR RX70 User s Manual MEMO...
Page 119: ...5 APPENDIX APPENDIX Error Messages and Countermeasures 118 Index 121 Revision History 124...
Page 122: ...Error Messages and Countermeasures 120 ZR RX70 User s Manual MEMO...
Page 125: ...Index ZR RX70 User s Manual 123 5 APPENDIX MEMO...