Logic Alarm Cable Connection and Functions
ZR-RX70 User’s Manual
2
CHECKS AND
PREPARATION
27
External trigger/sampling Specifications
Internal Equivalence Circuit
Alarm Output Specifications
Internal Equivalence Circuit and Example of Wiring Connection
Item
Description
Number of input channels
1
Input voltage range
0 to +24V max. (single-ended ground input)
Threshold level
Approx. +2.5V
Hysteresis
Approx. 0.5 V (+2.5 to +3 V)
Item
Description
Number of input channels
4
Output format
Open collector output
+5 V, 10 K
Ω
pull-up resistance
Contact capacity 5 V to 24 V, 100 mA or below