Setting Menus
64
ZR-RX70 User’s Manual
(2)-1 Sampling Interval
Specifies intervals used to capture data.
When the External Sampling is set, data is captured at 10µs as internal processing. Then, the input of an external
pulse allows the next internal processing data to be written into the memory.
Therefore, both capturing interval and input error become up to 10µs.
You can use the external sampling function to the extent that the input error can be ignored.
Capture destination Selections available
Internal RAM
10, 20, 50, 100, 200, 500us; 1, 2, 5, 10, 20, 50, 100, 200, 500ms; 1, 2, 5, 10, 20, 30s; 1min,
External
Internal flash memory 1, 2, 5, 10, 20, 50, 100, 200, 500ms; 1, 2, 5, 10, 20, 30s; 1min
USB memory
1, 2, 5, 10, 20, 50, 100, 200, 500ms; 1, 2, 5, 10, 20, 30s; 1min
• When the data capture destination is the internal flash memory or USB memory and the capture file format is Text
(CSV), sampling intervals from 1 to 5ms cannot be selected.
• When External is selected in the sampling interval, you cannot select External Input in the trigger setting. When
the External Input is already set, the trigger is changed to Off.
Important
This data
This data
10s
Input signal
External sampling
pulse
Internal processing
data capturing
interval
Data captured to
memory
Note
Summary of Contents for ZR-RX70
Page 1: ...User s Manual Portable Multi Logger ZR RX70 Cat No Z283 E1 02...
Page 12: ...10 ZR RX70 User s Manual MEMO...
Page 16: ...14 ZR RX70 User s Manual MEMO...
Page 108: ...WEB Server Function 106 ZR RX70 User s Manual MEMO...
Page 118: ...External Dimensions 116 ZR RX70 User s Manual MEMO...
Page 119: ...5 APPENDIX APPENDIX Error Messages and Countermeasures 118 Index 121 Revision History 124...
Page 122: ...Error Messages and Countermeasures 120 ZR RX70 User s Manual MEMO...
Page 125: ...Index ZR RX70 User s Manual 123 5 APPENDIX MEMO...