35
Run-Mode Measurement Process
Overall Judgement
How to get Overall Judgement Output (OR Output)
1.
Parallel OR Signal
OR signal’s (ON/OFF) indicates the total judgement (OK or NG).
Setting about Overall Judgement
Setting Reflection of Individual Results
MENU mode - [Setup] - [Add func] - [OR setting]
You can select which items results are reflected in the overall judgement that is
output to the OR signal of the parallel interface.
Measurement Item
ON (default) / OFF
Position correction
Calc./variable (Individual Data)
Calc./judge (Individual Judge)
Calc./alarm (Logging Monitor Alarm)
ON / OFF (default)
Able / Disable
MENU mode - [System] - [Output] - [Total jg. output] - [Parallel]
If selecting OFF, OR signal is disable.
ON (default) / OFF
Output Polarity
(OR, DO[0:15])
MENU mode - [System] - [Comm] - [Parallel] - [Polarity]
ON condition of OR and Individual Judgement
NG=ON: Signals turn ON when judgement is NG. (default )
OK=ON: Signals turn ON when judgement is OK.
OR output mode
MENU mode - [System] - [Comm] - [OR output]
One-shot: OR signals turns ON for specified time only when
ON condition is satisfied.
Level:
ON/OFF status is held until it next changes after OR signal
has been output. (default)
OR Output time
MENU mode - [System] - [Comm] - [OR output]
Output time of OR signal as a one-shot signal
Range:
0 to 255 ms (default: 0 ms)
Z01E-EN-01+ZFX+SettingGuide.book Seite 35 Mittwoch, 13. Februar 2008 2:38 14