16

How to Take Care of the Unit

How to Take Care of the Air Filter

Replace the air filter when it becomes dirty.

 Remove the nebulisation unit from the main 

device.

 Remove the fan cover by lifting it up.

 Remove the air filter case from inside the fan 

cover.

 Remove the air filter from the air filter case to 

replace with a new one.

Assemble the above mentioned parts in the 
reverse order.

How to Take Care of the Fan

Open the fan cover and remove the fan.
Clean the fan by washing it with water or detergent.

– If the medication is stuck to the fan installation 

axis, remove the medication.

Remove the fan cover by lifting it up after removing 
the nebulisation unit from the main device. The 
dirty fan cover can be washed with water or 
detergent.

In order to install the fan, position the fan to the fan 
installation axis securely as far as it goes.

How to Take Care of the Main Unit

Clean the dust on the main unit by wiping it with a 
lightly squeezed cloth after being moistened with 
water or alcohol.

After draining water from the water tank of the 
main unit through the draining hose, wipe off the 
leftover water with a clean and dry cloth.

3

4

1

2

General Advice

 • Do not wipe the main unit with volatile 

chemicals, such as benzine or thinners.

 • Do not damage the surface of the vibrator.

Nebulisation 
section

Nebulisation 
section

Fan installation 
axis

Fan

Fan cover

EN

NE-U17-E.indb   16

NE-U17-E.indb   16

14/12/2010   13:49:28

14/12/2010   13:49:28

Summary of Contents for Ultra Air NE-U17

Page 1: ...l professionals Only use medications as prescribed and instructed by your doctor Table of Contents Be sure to Read This Section Exemptions from Liabilities 1 Intended Use 2 Notes on Safety 3 About the...

Page 2: ...including the power source and installation environment 6 Trouble and or damage caused as a result of modifying and or improper repair of the product 7 Trouble and or damage caused by supreme forces s...

Page 3: ...who are unconscious or are not breathing spontaneously Environment This product is intended for use in a medical facility such as Hospital clinic and doctor s office Durable period Durable periods are...

Page 4: ...it for a long time be sure to clean and disinfect it Refer to Pages 15 to 17 Miscellaneous types of bacteria may propagate and you may get infected Clean and disinfect the medication cup medication cu...

Page 5: ...or damaged Do not fill the water tank with the liquid other than water tap water or distilled water The vibrator will deteriorate and may be damaged When parts are sterilised in an autoclave make sure...

Page 6: ...tank of the main unit Do not disinfect the unit by boiling in a microwave oven After cleaning and disinfecting the parts assemble them after they are completely dried Do not wipe the main unit with v...

Page 7: ...the Product 1 Easy to see large LCD displays air volume nebulisation rate set time remaining time and error location The back light illumination featured in the unit enables checking of the display i...

Page 8: ...IMER MIN WATER LEVEL MEDICATION CUP COVER FAN COVER TIMER POWER CONTINUOUS CONTINUOUS 1 2 3 START STOP Display of air volume The air volume is displayed in 11 levels Display of nebulisation rate The n...

Page 9: ...tion of the shortage of water Fan installation axis Fan Top of the main unit Back of the main unit START STOP Button Power cord Power Button Display of air volume Refer to the Description of the LCD D...

Page 10: ...If the water level in the water tank is lower than the water level mark a lower water level in water tank error will be displayed If the water temperature in the water tank is low the nebulisation rat...

Page 11: ...piece set 3 Turn on the Power Button After all items on the LCD display illuminate check that the ERROR display does not flash If the ERROR display flashes check the place where the error number is in...

Page 12: ...lation hose For inhalation using the accessories such as mouthpiece adjust the air volume and the nebulisation rate according to your preferences and the accessories being used The air volume 11 level...

Page 13: ...s the completion of time set by the timer and the unit stops operation When the operation ends 0 on the display illuminates and the display changes to the initial set time To terminate inhalation whil...

Page 14: ...the tube is not bent If the tube is bent the medication cannot be supplied If the tube is too long cut it to a proper length Attach the inhalation hose Set the timer to Continuous Turn the air volume...

Page 15: ...uct cover from the main unit Remove the bacterial filter cap of the long time nebulisation set Attach a bacterial filter to the long time nebulisation set Attach a connector to the bacterial filter At...

Page 16: ...the inside of the water tank by using alcohol Clean disinfect and immediately dry the medication cup cover medication cup holder medication cup inhalation hose mouthpiece set For the disinfection meth...

Page 17: ...t up after removing the nebulisation unit from the main device The dirty fan cover can be washed with water or detergent In order to install the fan position the fan to the fan installation axis secur...

Page 18: ...nnot be disinfected by boiling Typical disinfectants that can be used Disinfecting ethanol Glutaraldehyde Sterihyde 0 5 and other equivalent products Sodium hypochlorite Milton 0 1 Purax 0 1 and other...

Page 19: ...ilization by autoclave Refer to the Instruction Manual and the Notes on Handling of the sterilizer How to store the unit Store the parts after thoroughly drying them When storing the parts be careful...

Page 20: ...ug the power plug in the electric outlet correctly Refer to Page 10 Fill the water up to the specified water level mark in the operating chamber Refer to Page 9 Set the medication cup correctly Refer...

Page 21: ...Undersize Independently measured at SolAero Ltd Canada Dr John Dennis according to EN13544 1 2007 Measured by OMRON HEALTHCARE Co Ltd Notes Subject to technical modification without prior notice This...

Page 22: ...he unit and create a potentially unsafe situation Recommendation is to keep a minimum distance of 7 m Verify correct operation of the device in case the distance is shorter Further documentation in ac...

Page 23: ...hed with the fixing rubber string Enables the simultaneous inhalation by mouth and nose Material SEBS Packaging unit 3 sets Inhalation mask S set 4997207 0 Attached with the fixing rubber string Exclu...

Page 24: ...SUSAKA Co Ltd 1855 370 Kubo cho Matsusaka city Mie prefecture 515 8503 Japan Subsidiary OMRON HEALTHCARE UK LIMITED Opal Drive Fox Milne Milton Keynes MK15 0DG U K OMRON MEDIZINTECHNIK HANDELSGESELLSC...

Reviews: