How to Prepare the Unit before Inhalation

– Clean the parts before starting the inhalation.

[Parts that can be disinfected by boiling]

(For the disinfecting method, refer to Page 17.)

[Parts that cannot be disinfected by boiling]

(For the care and disinfection, refer to Page 17.)

Submerge these parts in a container filled with 
sufficient water and boil for 10 minutes.

Medication 
cup cover

Washer

Be sure to immediately dry the cleaned and disinfected 
parts, then store them in a dry place in order to prevent 
them from getting contaminated in the future.

1

Filling the cooling water.

1)  Remove the nebulisation section 

by rotating the medication cup 
cover-fixing lever clockwise.

2

1

water leve

l

2)  Fill the cooling water.

Fill the cooling water up to the “water 
level” mark that specifies the water level 
in the water tank. The proper amount is 
approximately 375 ml.

water le

vel

water level

–  If the water level in the water tank is lower than 

the “water level” mark, a “lower water level in water 
tank” error will be displayed.

–  If the water temperature in the water tank is low, 

the nebulisation rate may drop. (Recommended 
temperature is approximately 26°C (78.8 °F).

–  Replace water in the water tank each time the unit 

is used.

–  When the unit is used for a long time continuously, 

replace water at least every twenty-four hours.

  (For continuous use or long periods, refer to 

Page 13.)

Mouthpiece 
set

Air duct cover

Fan cover

Air filter case

Air filter

Fan

Medication cup

Medication cup holder

Warning

Inhalation hose (M) 

with a cuff

 How to remove the washer 

Place the medication cup cover 
on its side and insert a small item 
(such as a paper clip) into the 
hole of the cover.
Pinch and remove the washer 
gently.

 How to attach the washer 

Hold the medication cup cover and 
place the washer securely with 
your fingertips as illustrated in the 
figures on the left.
Make sure the washer is placed 
flat and not upside down.

NE-U17-E.indb   9

NE-U17-E.indb   9

14/12/2010   13:49:24

14/12/2010   13:49:24

Summary of Contents for Ultra Air NE-U17

Page 1: ...l professionals Only use medications as prescribed and instructed by your doctor Table of Contents Be sure to Read This Section Exemptions from Liabilities 1 Intended Use 2 Notes on Safety 3 About the...

Page 2: ...including the power source and installation environment 6 Trouble and or damage caused as a result of modifying and or improper repair of the product 7 Trouble and or damage caused by supreme forces s...

Page 3: ...who are unconscious or are not breathing spontaneously Environment This product is intended for use in a medical facility such as Hospital clinic and doctor s office Durable period Durable periods are...

Page 4: ...it for a long time be sure to clean and disinfect it Refer to Pages 15 to 17 Miscellaneous types of bacteria may propagate and you may get infected Clean and disinfect the medication cup medication cu...

Page 5: ...or damaged Do not fill the water tank with the liquid other than water tap water or distilled water The vibrator will deteriorate and may be damaged When parts are sterilised in an autoclave make sure...

Page 6: ...tank of the main unit Do not disinfect the unit by boiling in a microwave oven After cleaning and disinfecting the parts assemble them after they are completely dried Do not wipe the main unit with v...

Page 7: ...the Product 1 Easy to see large LCD displays air volume nebulisation rate set time remaining time and error location The back light illumination featured in the unit enables checking of the display i...

Page 8: ...IMER MIN WATER LEVEL MEDICATION CUP COVER FAN COVER TIMER POWER CONTINUOUS CONTINUOUS 1 2 3 START STOP Display of air volume The air volume is displayed in 11 levels Display of nebulisation rate The n...

Page 9: ...tion of the shortage of water Fan installation axis Fan Top of the main unit Back of the main unit START STOP Button Power cord Power Button Display of air volume Refer to the Description of the LCD D...

Page 10: ...If the water level in the water tank is lower than the water level mark a lower water level in water tank error will be displayed If the water temperature in the water tank is low the nebulisation rat...

Page 11: ...piece set 3 Turn on the Power Button After all items on the LCD display illuminate check that the ERROR display does not flash If the ERROR display flashes check the place where the error number is in...

Page 12: ...lation hose For inhalation using the accessories such as mouthpiece adjust the air volume and the nebulisation rate according to your preferences and the accessories being used The air volume 11 level...

Page 13: ...s the completion of time set by the timer and the unit stops operation When the operation ends 0 on the display illuminates and the display changes to the initial set time To terminate inhalation whil...

Page 14: ...the tube is not bent If the tube is bent the medication cannot be supplied If the tube is too long cut it to a proper length Attach the inhalation hose Set the timer to Continuous Turn the air volume...

Page 15: ...uct cover from the main unit Remove the bacterial filter cap of the long time nebulisation set Attach a bacterial filter to the long time nebulisation set Attach a connector to the bacterial filter At...

Page 16: ...the inside of the water tank by using alcohol Clean disinfect and immediately dry the medication cup cover medication cup holder medication cup inhalation hose mouthpiece set For the disinfection meth...

Page 17: ...t up after removing the nebulisation unit from the main device The dirty fan cover can be washed with water or detergent In order to install the fan position the fan to the fan installation axis secur...

Page 18: ...nnot be disinfected by boiling Typical disinfectants that can be used Disinfecting ethanol Glutaraldehyde Sterihyde 0 5 and other equivalent products Sodium hypochlorite Milton 0 1 Purax 0 1 and other...

Page 19: ...ilization by autoclave Refer to the Instruction Manual and the Notes on Handling of the sterilizer How to store the unit Store the parts after thoroughly drying them When storing the parts be careful...

Page 20: ...ug the power plug in the electric outlet correctly Refer to Page 10 Fill the water up to the specified water level mark in the operating chamber Refer to Page 9 Set the medication cup correctly Refer...

Page 21: ...Undersize Independently measured at SolAero Ltd Canada Dr John Dennis according to EN13544 1 2007 Measured by OMRON HEALTHCARE Co Ltd Notes Subject to technical modification without prior notice This...

Page 22: ...he unit and create a potentially unsafe situation Recommendation is to keep a minimum distance of 7 m Verify correct operation of the device in case the distance is shorter Further documentation in ac...

Page 23: ...hed with the fixing rubber string Enables the simultaneous inhalation by mouth and nose Material SEBS Packaging unit 3 sets Inhalation mask S set 4997207 0 Attached with the fixing rubber string Exclu...

Page 24: ...SUSAKA Co Ltd 1855 370 Kubo cho Matsusaka city Mie prefecture 515 8503 Japan Subsidiary OMRON HEALTHCARE UK LIMITED Opal Drive Fox Milne Milton Keynes MK15 0DG U K OMRON MEDIZINTECHNIK HANDELSGESELLSC...

Reviews: