How to Inhale (Using Optional Parts)

1

Using the unit for a long time

By using the optional “long time nebulisation set (4997217-7)” and the “supply medication set (4997206-1)”, 
the unit can nebulise continuously for a long time.

 Remove the medication cup cover of the main unit.  Attach an optional long time nebulisation set in place 

of the medication cup cover and assemble the unit according to the procedure shown on Page 10.

  At this time, check that the air duct is not blocked by the air duct cap.

 Fill medication into the supply medication bottle.

 Tightly close the supply medication bottle cap.  At this time, check that the washer is correctly set inside 

the cap.

 Remove the cap of the supply medication pipe of the long time nebulisation set and attach the tube of the 

supply medication set.

 Hang the supply medication bottle on a proper place, higher than the device. 

  At this time, check that the tube is not bent.  If the tube is bent, the medication cannot be supplied.  If the 

tube is too long, cut it to a proper length.

  Attach the inhalation hose. 

 Set the timer to “Continuous”.

 Turn the air volume adjustment knob to the far right (the position of 10). 

 Push the START/STOP Button (

) and start inhalation.

If you use the unit for humidification in combination with oxygen therapy, connect the equipment to the 
oxygen supply pipe of the “long time nebulisation set” (optional, 4997217-7). Please refer to the illustration 
on the next page.

In order to stop the air from the fan on the main unit during the oxygen therapy, set the “air-flow volume” 
adjustment knob to “0”, and attach an air duct cap to the air duct of the long time nebulisation set.

The operating and inhalation methods are the same as those of the normal operation.

Using the unit for (oxygen) humidification

2

Cap

Supply 
medication 
pipe

Air duct 
cap

Air duct

Supply 
medication set 
(4997206-1)

Supply 
medication 
bottle

Long time nebulisation 
set (4997217-7)

Cap

Tube

NE-U17-E.indb   13

NE-U17-E.indb   13

14/12/2010   13:49:27

14/12/2010   13:49:27

Summary of Contents for Ultra Air NE-U17

Page 1: ...l professionals Only use medications as prescribed and instructed by your doctor Table of Contents Be sure to Read This Section Exemptions from Liabilities 1 Intended Use 2 Notes on Safety 3 About the...

Page 2: ...including the power source and installation environment 6 Trouble and or damage caused as a result of modifying and or improper repair of the product 7 Trouble and or damage caused by supreme forces s...

Page 3: ...who are unconscious or are not breathing spontaneously Environment This product is intended for use in a medical facility such as Hospital clinic and doctor s office Durable period Durable periods are...

Page 4: ...it for a long time be sure to clean and disinfect it Refer to Pages 15 to 17 Miscellaneous types of bacteria may propagate and you may get infected Clean and disinfect the medication cup medication cu...

Page 5: ...or damaged Do not fill the water tank with the liquid other than water tap water or distilled water The vibrator will deteriorate and may be damaged When parts are sterilised in an autoclave make sure...

Page 6: ...tank of the main unit Do not disinfect the unit by boiling in a microwave oven After cleaning and disinfecting the parts assemble them after they are completely dried Do not wipe the main unit with v...

Page 7: ...the Product 1 Easy to see large LCD displays air volume nebulisation rate set time remaining time and error location The back light illumination featured in the unit enables checking of the display i...

Page 8: ...IMER MIN WATER LEVEL MEDICATION CUP COVER FAN COVER TIMER POWER CONTINUOUS CONTINUOUS 1 2 3 START STOP Display of air volume The air volume is displayed in 11 levels Display of nebulisation rate The n...

Page 9: ...tion of the shortage of water Fan installation axis Fan Top of the main unit Back of the main unit START STOP Button Power cord Power Button Display of air volume Refer to the Description of the LCD D...

Page 10: ...If the water level in the water tank is lower than the water level mark a lower water level in water tank error will be displayed If the water temperature in the water tank is low the nebulisation rat...

Page 11: ...piece set 3 Turn on the Power Button After all items on the LCD display illuminate check that the ERROR display does not flash If the ERROR display flashes check the place where the error number is in...

Page 12: ...lation hose For inhalation using the accessories such as mouthpiece adjust the air volume and the nebulisation rate according to your preferences and the accessories being used The air volume 11 level...

Page 13: ...s the completion of time set by the timer and the unit stops operation When the operation ends 0 on the display illuminates and the display changes to the initial set time To terminate inhalation whil...

Page 14: ...the tube is not bent If the tube is bent the medication cannot be supplied If the tube is too long cut it to a proper length Attach the inhalation hose Set the timer to Continuous Turn the air volume...

Page 15: ...uct cover from the main unit Remove the bacterial filter cap of the long time nebulisation set Attach a bacterial filter to the long time nebulisation set Attach a connector to the bacterial filter At...

Page 16: ...the inside of the water tank by using alcohol Clean disinfect and immediately dry the medication cup cover medication cup holder medication cup inhalation hose mouthpiece set For the disinfection meth...

Page 17: ...t up after removing the nebulisation unit from the main device The dirty fan cover can be washed with water or detergent In order to install the fan position the fan to the fan installation axis secur...

Page 18: ...nnot be disinfected by boiling Typical disinfectants that can be used Disinfecting ethanol Glutaraldehyde Sterihyde 0 5 and other equivalent products Sodium hypochlorite Milton 0 1 Purax 0 1 and other...

Page 19: ...ilization by autoclave Refer to the Instruction Manual and the Notes on Handling of the sterilizer How to store the unit Store the parts after thoroughly drying them When storing the parts be careful...

Page 20: ...ug the power plug in the electric outlet correctly Refer to Page 10 Fill the water up to the specified water level mark in the operating chamber Refer to Page 9 Set the medication cup correctly Refer...

Page 21: ...Undersize Independently measured at SolAero Ltd Canada Dr John Dennis according to EN13544 1 2007 Measured by OMRON HEALTHCARE Co Ltd Notes Subject to technical modification without prior notice This...

Page 22: ...he unit and create a potentially unsafe situation Recommendation is to keep a minimum distance of 7 m Verify correct operation of the device in case the distance is shorter Further documentation in ac...

Page 23: ...hed with the fixing rubber string Enables the simultaneous inhalation by mouth and nose Material SEBS Packaging unit 3 sets Inhalation mask S set 4997207 0 Attached with the fixing rubber string Exclu...

Page 24: ...SUSAKA Co Ltd 1855 370 Kubo cho Matsusaka city Mie prefecture 515 8503 Japan Subsidiary OMRON HEALTHCARE UK LIMITED Opal Drive Fox Milne Milton Keynes MK15 0DG U K OMRON MEDIZINTECHNIK HANDELSGESELLSC...

Reviews: