Troubleshooting

Trouble

The power 
lamp does not 
illuminate.

The unit does not 
nebulise.

The nebulisation 
rate is low.

The nebulisation 
power is lowered.

Nebulisation is 
unstable.

Where to Inspect

Is the power plug securely plugged into the 
electric outlet?

Is 

 of ERROR display flashing due to 

insufficient amount of water in the operating 
chamber?

Is the medication cup correctly set to the 
main unit?

Is the fan cover open?

Is the fan set?

Is the amount of medication in the medication 
cup too much?

Is the inhalation hose bent and the medication 
pooled in the hose? 

Is the amount of medication in the medication 
cup too small?

Is the lower portion of medication cup 
damaged or deformed?

Is the amount of medication in the medication 
cup much?

Is the value of nebulisation rate adjustment 
knob or the air volume adjustment knob set 
too low?

Is the room temperature and/or water 
temperature too low?

Is the medication cup dirty?

Is the amount of medication exceeding 
150 ml?

Are the air volume and the nebulisation rate 
set to the maximum level when the amount of 
medication is 150 ml?

– If the unit does not operate normally after taking the above-mentioned 

measures, do not touch the internal mechanism and consult the store where 
you purchased the unit or the nearest OMRON dealer.

How to Correct

Plug the power plug in the electric outlet 
correctly.   

(Refer to Page 10.)

Fill the water up to the specified water level 
mark in the operating chamber.
 

(Refer to Page 9.)

Set the medication cup correctly.
 

(Refer to Page 10.)

Close the fan cover correctly.

Set the fan properly.

Reduce the amount of medication to less than 
150 ml. 

(Refer to Page 10.)

Adjust the bent hose correctly and drain the 
pooled medication.  

(Refer to Page 12.)

Increase the amount of medication to more 
than 5 ml.

Replace the medication cup with a new one.

Reduce the amount of medication to less than 
150 ml. 

(Refer to Page 10.)

Turn the knob to the right and adjust the value.  
 

(Refer to Page 11.)

Operate the unit without medication for about 
4 minutes, then use the unit.

After cleaning the medication cup with the 
detergent, wash off the detergent sufficiently 
under running water before using the cup.

Reduce the amount of medication to less 
than 150 ml. 

(Refer to Page 10.)

Adjust the nebulisation rate. (Refer to Page 11.)

Trouble

Where to Inspect

How to Correct

 flashes.

 flashes.

 flashes.

E1 is displayed.

The water level in the operating 
chamber has been lowered.

The medication cup cover is open.

The fan cover is open.

The fan inside the main unit has 
stopped.

Fill in the water up to the water level 
of the operating chamber.

Securely close the medication cup 
cover.

Securely close the fan cover.

Contact the nearest OMRON dealer.

WATER LEVEL
MEDICATION CUP

COVER
FAN COVER

1

2

3

NE-U17-E.indb   19

NE-U17-E.indb   19

14/12/2010   13:49:29

14/12/2010   13:49:29

Summary of Contents for Ultra Air NE-U17

Page 1: ...l professionals Only use medications as prescribed and instructed by your doctor Table of Contents Be sure to Read This Section Exemptions from Liabilities 1 Intended Use 2 Notes on Safety 3 About the...

Page 2: ...including the power source and installation environment 6 Trouble and or damage caused as a result of modifying and or improper repair of the product 7 Trouble and or damage caused by supreme forces s...

Page 3: ...who are unconscious or are not breathing spontaneously Environment This product is intended for use in a medical facility such as Hospital clinic and doctor s office Durable period Durable periods are...

Page 4: ...it for a long time be sure to clean and disinfect it Refer to Pages 15 to 17 Miscellaneous types of bacteria may propagate and you may get infected Clean and disinfect the medication cup medication cu...

Page 5: ...or damaged Do not fill the water tank with the liquid other than water tap water or distilled water The vibrator will deteriorate and may be damaged When parts are sterilised in an autoclave make sure...

Page 6: ...tank of the main unit Do not disinfect the unit by boiling in a microwave oven After cleaning and disinfecting the parts assemble them after they are completely dried Do not wipe the main unit with v...

Page 7: ...the Product 1 Easy to see large LCD displays air volume nebulisation rate set time remaining time and error location The back light illumination featured in the unit enables checking of the display i...

Page 8: ...IMER MIN WATER LEVEL MEDICATION CUP COVER FAN COVER TIMER POWER CONTINUOUS CONTINUOUS 1 2 3 START STOP Display of air volume The air volume is displayed in 11 levels Display of nebulisation rate The n...

Page 9: ...tion of the shortage of water Fan installation axis Fan Top of the main unit Back of the main unit START STOP Button Power cord Power Button Display of air volume Refer to the Description of the LCD D...

Page 10: ...If the water level in the water tank is lower than the water level mark a lower water level in water tank error will be displayed If the water temperature in the water tank is low the nebulisation rat...

Page 11: ...piece set 3 Turn on the Power Button After all items on the LCD display illuminate check that the ERROR display does not flash If the ERROR display flashes check the place where the error number is in...

Page 12: ...lation hose For inhalation using the accessories such as mouthpiece adjust the air volume and the nebulisation rate according to your preferences and the accessories being used The air volume 11 level...

Page 13: ...s the completion of time set by the timer and the unit stops operation When the operation ends 0 on the display illuminates and the display changes to the initial set time To terminate inhalation whil...

Page 14: ...the tube is not bent If the tube is bent the medication cannot be supplied If the tube is too long cut it to a proper length Attach the inhalation hose Set the timer to Continuous Turn the air volume...

Page 15: ...uct cover from the main unit Remove the bacterial filter cap of the long time nebulisation set Attach a bacterial filter to the long time nebulisation set Attach a connector to the bacterial filter At...

Page 16: ...the inside of the water tank by using alcohol Clean disinfect and immediately dry the medication cup cover medication cup holder medication cup inhalation hose mouthpiece set For the disinfection meth...

Page 17: ...t up after removing the nebulisation unit from the main device The dirty fan cover can be washed with water or detergent In order to install the fan position the fan to the fan installation axis secur...

Page 18: ...nnot be disinfected by boiling Typical disinfectants that can be used Disinfecting ethanol Glutaraldehyde Sterihyde 0 5 and other equivalent products Sodium hypochlorite Milton 0 1 Purax 0 1 and other...

Page 19: ...ilization by autoclave Refer to the Instruction Manual and the Notes on Handling of the sterilizer How to store the unit Store the parts after thoroughly drying them When storing the parts be careful...

Page 20: ...ug the power plug in the electric outlet correctly Refer to Page 10 Fill the water up to the specified water level mark in the operating chamber Refer to Page 9 Set the medication cup correctly Refer...

Page 21: ...Undersize Independently measured at SolAero Ltd Canada Dr John Dennis according to EN13544 1 2007 Measured by OMRON HEALTHCARE Co Ltd Notes Subject to technical modification without prior notice This...

Page 22: ...he unit and create a potentially unsafe situation Recommendation is to keep a minimum distance of 7 m Verify correct operation of the device in case the distance is shorter Further documentation in ac...

Page 23: ...hed with the fixing rubber string Enables the simultaneous inhalation by mouth and nose Material SEBS Packaging unit 3 sets Inhalation mask S set 4997207 0 Attached with the fixing rubber string Exclu...

Page 24: ...SUSAKA Co Ltd 1855 370 Kubo cho Matsusaka city Mie prefecture 515 8503 Japan Subsidiary OMRON HEALTHCARE UK LIMITED Opal Drive Fox Milne Milton Keynes MK15 0DG U K OMRON MEDIZINTECHNIK HANDELSGESELLSC...

Reviews: