3-5 Adjusting the Touch Panel and Display
3-5-1 Calibrating the Touch Panel
3-18
3
In
st
alling t
h
e NSA-
ser
ies I
n
dust
rial PC and Connect
ing Per
ipher
al Devices
3-5 Adjusting the Touch Panel and
Display
3-5-1 Calibrating the Touch Panel
This section describes the procedure for calibrating the touch panel.
1. Calibrate the touch panel whenever it deviates markedly from its coordinate.
Connect a PS/2 keyboard and mouse or a USB keyboard and mouse to the
NSA-series Industrial PC before calibrating the touch panel. Make sure the
keyboard and mouse are properly connected.
Note
Shut down the NSA-series Industrial PC operating system and turn OFF the NSA-series
Industrial PC before you connect the PS/2 keyboard and mouse.
2. Start up the NSA-series Industrial PC and select Start – Programs – UPDD –
Settings from the bottom left corner of the screen to launch the settings
program.
3. Click the Calibration Tab on the Pointer Device Properties Menu, then click
the Calibrate Button.
Calibrate Button
Make sure the EEPROM
Calibration box is selected.
Settings