Calibration (0 to 5 V or 1 to 5 V)
1, 2, 3...
1. First, the
sure
alarm display will appear. This will not appear once field cal-
ibration is performed.
Set Value LED Display Model
Basic Model
2. Press the Mode Key to display
1U s
. The teaching indicator will be lit.
Set Value LED Display Model
Basic Model
3. Press the RESET/TEACH Key to display the calibration value setting. Then
set the output of STV to 5 V. The teaching indicator will flash. After the value
stabilizes (i.e., the fluctuation of the value is less than several digits), press
the Mode Key to tentatively register the calibration data.
Set Value LED Display Model
Basic Model
4. The display will change to
1U 0
for 0-V calibration. The teaching indicator will
be lit.
Set Value LED Display Model
Basic Model
5. Press the RESET/TEACH Key to display the calibration value setting. Then
set the output of STV to 0 V. The teaching indicator will flash. After the value
stabilizes (i.e., the fluctuation of the value is less than several digits), press
the Mode Key to tentatively register the calibration data.
Set Value LED Display Model
Basic Model
6. The display will change to
str
for confirmation of the settings.
Set Value LED Display Model
Basic Model
Maintenance Mode