3. Press the RESET/TEACH Key to display the calibration value setting. Then
set Dial 6 to 300
Ω
. The teaching indicator will flash. After the value stabilizes
(i.e., the fluctuation of the value is less than several digits), press the Mode
Key to tentatively register the calibration data.
Set Value LED Display Model
Basic Model
4. The display will change to
p 0
for 0-
Ω
calibration. The teaching indicator will
be lit.
Set Value LED Display Model
Basic Model
Short-circuit terminals 2 to 4.
5. Press the RESET/TEACH Key to display the calibration value setting. The
teaching indicator will flash. After the value stabilizes (i.e., the fluctuation of
the value is less than several digits), press the Mode Key to tentatively regis-
ter the calibration data.
Set Value LED Display Model
Basic Model
6. Calibrate the B-to-B’ input next. The display will change to
pb 10
for 10-
Ω
calibration. Change the wiring as shown below.
Dial 6
Make the wiring distance between the Dial 6 and terminals 2 and 4 as short
as possible. Short-circuit terminals 2 and 3 to each other.
Set Value LED Display Model
Basic Model
Change wiring.
7. Press the RESET/TEACH Key to display the calibration value setting. Then
set the Dial 6 to 10
Ω
. The teaching indicator will flash. After the value stabi-
lizes (i.e., the fluctuation of the value is less than several digits), press the
Mode Key to tentatively register the calibration data.
Set Value LED Display Model
Basic Model
Maintenance Mode
Section 5-3