92
5. Press the RESET/TEACH Key to display the calibration value setting. Then
set the output of STV to 0 V. The teaching indicator will flash. After the value
stabilizes (i.e., the fluctuation of the value is less than several digits), press
the Mode Key to tentatively register the calibration data.
Set Value LED Display Model
Basic Model
6. The display will change to
str
for confirmation of the settings.
Set Value LED Display Model
Basic Model
7. Press the Shift Key to select whether or not to accept the settings.
Set Value LED Display Model
Basic Model
8. Press the Up Key to select
yes
.
9. Press the Mode Key to overwrite all calibration data and to put the K3NH into
RUN mode. The calibration data will not be overwritten if
no
is selected.
Set Value LED Display Model
Basic Model
RUN mode
Maintenance Mode
Section 5-3