5.0 WARNINGS
General Warnings
No modification of this equipment by the user is allowed. Modification will void the warranty.
BEACON laser system is designed to be used only with OmniGuide fibers. Do not
use fibers which are not approved for BEACON laser system.
Do not use laser accessories which are not approved for the BEACON laser system. Examine
accessories for any sign of damage before use. Do not use damaged accessories.
If this equipment is modified, appropriate inspection and testing must be
conducted to ensure continued safe use of the equipment
To avoid risk of electric shock, this equipment must only be connected to supply mains with protective earth
BEACON Laser System weights more than 150 lbs and may cause injury if proper care is not
used when moving it. The system should always be moved slowly and carefully
Warnings for Laser Surgical Procedures
The BEACON Advanced Energy Laser System is a fDA Class 4 Laser device that produces invisible beams of high-
energy infrared radiation at 10.6 microns wavelength (Class 4). Articulating arm has also an output of the laser beam
at the visible wavelength of 635 nm (Class 2M). Improper use of the Laser could result in serious personal injury.
Proper electrical and Laser safety training and strict observation of all safety precautions as specified in the Laser
Instructions for Use or Operations Manual are required to operate the device and/or provide its maintenance.
Laser surgical procedures should be performed only by individuals adequately trained and familiar
with such surgical techniques and hazards associated with it. Consult medical literature regarding
techniques, complications, and hazards prior to performing these procedures.
All personnel in the immediate area of laser use must wear eye protection specifically rated for CO
2
Lasers (OD 4+). Failure to do so may result in serious and permanent damage to the eyes.
Laser beam coming out of the articulating arm is a non-divergent that should not be activated
without an accessory attached to the articulating arm. Laser beam should not be aimed at the
walls, floors or other object as it will create burn marks -even from a distance.
Before instruments and accessories from different manufacturers are employed in a procedure, the user must
verify compatibility with the OmniGuide System. If in doubt, contact OmniGuide for assistance.
The laser should be activated only when the laser treatment site is clearly observable and the fiber tip is directed at it.
A fire extinguisher must be available in the operating room when the laser is being
utilized. Saline solution or water should also be readily available.
Flammable prep solutions (e.g. alcohol-based prep solutions) must not be used during Laser procedures.
Follow established protocols for laser surgery to minimize the risk of airway fires.
A smoke evacuation system must be utilized to remove surgical plume.
Do not submerge any portion of The BEACON Advanced Energy Laser System in fluid.