L-1670 Rev 0
49
© OmniGuide, Inc.
ID
Error
Condition
Displayed text
A-12
Parameter
verification error
One or more LC parameters has changed
from its expected value (condition
present for more than 500ms).
System error. Please restart the laser.
If the problem persists, call OmniGuide
Service at 888-666-4484 ext 3.
GAS SYSTEM
A-15
Gas Overflow error
Gas flow exceeds max specified in Gas Flow
Critical Overrflow deviation while in READY state
Helium gas overflow error. Please check
fiber, filter and gas connections.
A-16
Gas Underflow error
Gas flow below minimum specified in Gas Flow
Critical Underflow deviation while in READY state.
Helium gas underflow error. Please check
fiber, filter and gas connections.
If problem persists, replace with a new OmniGuide
fiber or call OmniGuide Service at 888-666-4484 ext 3.
A-17
MfC Communication
failure
No communication with MfC when
operating in fiber modality
flow controller error. Please re-start the laser system.
If the problem persists, call OmniGuide
Service at 888-666-4484 ext 3.
A-18
Tank input pressure out of
range (one or both tanks)
Laser Controller reports 0 (zero) value for
tank pressure of one or both tanks (below
0 converted value per SW-034)
flow controller error. Please re-start the laser system.
If the problem persists, call OmniGuide
Service at 888-666-4484 ext 3.
A-19
MfC failed to tare
during POST
During POST, the tare function of the MfC
did not correctly zero the device.
flow controller error. Please re-start the laser system.
If the problem persists, call OmniGuide
Service at 888-666-4484 ext 3.
fOOT PEDAL
A-20
foot Pedal Malfunction
LC asserts fault bit – foot pedal failure
foot pedal error. Please reconnect
or replace the foot pedal.
If the problem persists, call OmniGuide
Service at 888-666-4484 ext 3.
SYSTEM
A-25
System Watchdog
Timeout
System watchdog times out
Software error. Please re-start the laser system.
If the problem persists, call OmniGuide
Service at 888-666-4484 ext 3.
A-26
Thread Watchdog
Timeout
Thread watchdog times out
Software error. Please re-start the laser system.
If the problem persists, call OmniGuide
Service at 888-666-4484 ext 3.
A-27
System Clock failure
POST failure checking clock accuracy
Software error. Please re-start the laser system.
If the problem persists, call OmniGuide
Service at 888-666-4484 ext 3.
A-28
EMO Pressed
Laser Controller reports that EMO (E-Stop) is pressed
EMO (Emergency Machine Off) button pressed.
Disengage EMO button to continue.
A-29
Door Interlock Active
Laser Controller reports that Door
Interlock (Remote E-Stop) is active
Door interlock active.
Check interlock connection or close door.
A-30
Laser Enclosure Opened
Laser enclosure is opened while in READY mode
Back panel enclosure is open.
Please close back panel to continue.
A-31
Software failure:
Data Corruption
Verification of data CRC or other
data validation method fails
Software error. Please re-start the laser system.
If the problem persists, call OmniGuide
Service at 888-666-4484 ext 3.
A-32
Software failure: Other
Software detects a memory leak, resource
failure, or some such failure
Software error. Please re-start the laser system.
If the problem persists, call OmniGuide
Service at 888-666-4484 ext 3
ENVIRONMENTAL
A-35
Humidity Too High
LC asserts Humidity fault bit and
Humidity reading is above 95%.
Ambient humidity too high to operate the laser.
If the problem persists, call OmniGuide
Service at 888-666-4484 ext 3.