Make sure that the mains ground connection is made. The mains cable is connected
to the rear of the laser into the socket with the following characteristics:
Input voltage: 120-240 V, 50/ 60 Hertz Ground connection required Leakage current
trip: less 500 μA Fuses 120 V /15A (North America), 240 V / 10A (Europe)
The mains socket should be properly fixed to the wall; it is not allowed to
use an extension cable between the Laser and the wall outlet.
11.1.1 Fuse Specification for the Laser
Different input voltages require different fuses.
240 V 10A
120 V 15A
Compressed Gas Requirement
The OmniGuide laser system is designed to be used with medical-grade Helium gas, delivered to the
fiber at a pressure that depends on the type of fiber and the application. Please see fiber IFU for details.
Adjustments of the setting will be automatically limited by the settings on the RFID tag of the particular fiber.
The gas flow is established by connecting the gas hose (connected to the system) to a Helium
source (provided by the user). The gas hose has a CGA-930 fitting to facilitate this attachment. The
gas outlet on the top of the laser has a 1/8 “ quick disconnect port. The gas hose supplied with the
system connects the gas outlet at the front of the laser to the fiber adapter. A Gas Filter Unit (P/N
ACC-GFU-100-1) between the gas hose and the adapter is required for sterile procedures.
Door Interlock
The Laser should be used only in a room which is clearly marked as laser room and could be
interlocked through the laser console. To allow the door to be wired into the laser, the laser console
has an interlock connector at the base of the housing next to the foot pedal (Figure 7-4).
The door interlock is a switch in normal open condition. The switch needs to be in closed condition
for the laser to operate. When the door is closed or the interlock defeated the laser can be operated.
If the switch is opened during laser operation, the laser will show an error condition and stop.
When door interlock is not mandated, a connector needs to be in place to
defeat the door interlock in order for the laser to be started. A door interlock
plug has to be used to allow laser operation (P/N 3-0142-015-31-00).
Foot Pedal Connector
The foot pedal is directly attached to the base of the laser (Figure 7-4). The foot pedal can be removed
by pulling the connector from the rear of the laser. The foot pedal is shielded with a protection cover.