L-1670 Rev 0
50
© OmniGuide, Inc.
ID
Error
Condition
Displayed text
A-36
Ambient Temperature
Too High
LC asserts Temperature fault bit and Temperature
reading is above the max operational temperature.
Ambient temperature too high to operate the laser.
If the problem persists, call OmniGuide
Service at 888-666-4484 ext 3.
A-37
Ambient Temperature
Too Low
LC asserts Temperature fault bit and Temperature
reading is below the min operational temperature.
Ambient temperature too low to operate the laser.
If the problem persists, call OmniGuide
Service at 888-666-4484 ext 3.
A-38
Moisture
Condensation Risk
LC asserts Humidity fault bit and humidity is
over 85% and temperature is over 25 C.
Ambient conditions present a high
risk of moisture condensation.
If the problem persists, call OmniGuide
Service at 888-666-4484 ext 3.
LASER TUBE OR POWER SUPPLY
A-40
Laser Tube Temperature
Too High
LC HW Status:
RC1 = 0;
RC2 = 0;
Voltage OK = 1;
Laser tube error. Call OmniGuide
Service at 888-666-4484 ext 3.
A-41
Power Supply failure
LC HW Status:
RC1 = 0;
RC2 = 0;
Laser Voltage OK = 0 (> one second
after RC1 & RC2 = 0)
Laser tube error. Call OmniGuide
Service at 888-666-4484 ext 3.
A-42
SWR fault
LC HW Status:
RC1 = 0;
RC2 = 0;
Voltage OK = 1;
Laser OK = 0;
Laser tube error. Call OmniGuide
Service at 888-666-4484 ext 3.
A-43
Power Supply
Did Not Start
During POST, the HW Control status relay
contacts indicate that the laser power supply
is not enabled (RC1 != 0 or RC2 != 0)
Laser tube error. Call OmniGuide
Service at 888-666-4484 ext 3.
A-44
Power Supply failure
During operation, the HW Control status relay
contacts indicate that the laser power supply
has been disabled (RC1 != 0 or RC2 != 0)
Laser tube error. Call OmniGuide
Service at 888-666-4484 ext 3.
LASER POWER
A-45
Tube Current Too High
Tube current higher than specified limit (20%
above nominal) when laser is firing.
Laser tube error. Call OmniGuide
Service at 888-666-4484 ext 3.
A-46
Tube Current Too Low
Tube current lower than specified limit (20%
below nominal) when laser is firing.
Laser tube error. Call OmniGuide
Service at 888-666-4484 ext 3.
A-47
Tube Current
Sensor Broken
Tube current is 0
Laser tube error. Call OmniGuide
Service at 888-666-4484 ext 3.
A-48
Output Power Too High
Pre-Ready power check indicates power and tube
current exceed nominal value by more than 20%
Laser tube error. Call OmniGuide
Service at 888-666-4484 ext 3.
A-49
Output Power Too Low
Pre-Ready power check indicates power and tube
current below nominal value by more than 20%
Laser tube error. Call OmniGuide
Service at 888-666-4484 ext 3.
POWER DETECTOR
A-50
Power Detector Too Low
Power detector signal is more than 20% below nominal
when tube current is valid for specified power
Power monitoring error. Call OmniGuide
Service at 888-666-4484 ext 3.
A-51
Power Detector Too high
Power detector signal is more than 20% above
nominal when tube current is valid for specified power
Power monitoring error. Call OmniGuide
Service at 888-666-4484 ext 3.
A-52
Power Detector Failure (1)
Power Detector fail fault bit asserted.
Power monitoring error. Call OmniGuide
Service at 888-666-4484 ext 3.