9.0 TRANSPORTATION AND STORAGE
If the system has been transported or stored at temperatures below 10°C (50 F), unpack the device and
leave it at normal room temperatures at least for half a day so that the laser reaches room temperature.
Store at temperatures: -30°C to 70°C
Pressure: 525 hPa to 1060 hPa
Humidity: 0-95%
Indicates upper side. Keep upright.
fragile – handle with care.
Keep dry
for extended storage, The BEACON Advanced Energy Laser System is
should be stored in dry environment at normal room temperatures.
Care should always be taken when moving the The BEACON Advanced Energy Laser System
system. Before moving the laser, it is recommended to disconnect the footswitch from the laser
and attach it to the back panel. A handle located on the front panel allows easy movement of
the system, but take special care when manoeuvring over thresholds, elevator doors, ramps and
other uneven or sloping floor surfaces. A severe physical shock could cause the alignment of the
laser or the articulated arm to be disturbed resulting in personal injury or physical damage.
The BEACON Advanced Energy Laser System is not designed for mobile use. If it becomes necessary
to relocate the laser, contact OMNIGUIDE Customer Support or your distributor for details. failure
to do so may result in personal injuries or damage to the system and may void any warranty.