Accuracy of the Analog DC Input
Note:
Exceeding the specified input values can damage the measurement
inputs!
Table 6-29:
DC measurement input
Table 6-30:
DC voltage measurement
input
DC Measurement Input IDC
Measurement ranges
0 ... ±1 mA
0 ... ±20 mA
Max. input current
600 mA
Accuracy
Typical error
< 0.003 % of rg.
1
1
rg. = range, whereat
n
% of rg. means:
n
% of upper range value.
Guaranteed error
< 0.02 % of rg.
Input impedance
Approx. 15
Ω
Connection
4 mm/0.16 " banana connectors
Insulation
Insulation to all other front panel connections.
Reinforced insulation from all SELV interfaces and
from power supply.
Not galvanically isolated from V
DC
.
DC Voltage Measurement Input VDC
Measurement range
0…± 10 V
Max. input voltage
± 11 V
Input impedance
1 M
Ω
Max. input current
± 90 mA
Accuracy
Typical error
< 0.003 % of rg.
Guaranteed error
< 0.02 % of rg.
Connection
4 mm/0.16 " banana connectors
Insulation
Not galvanically isolated from I
DC