CMC 356 Reference Manual
56
6.3.6 Low-Level
Binary
Outputs ("ext. Interf.")
The SELV interface connector "ext. Interf." holds four additional transistor
binary outputs (Bin. out 11 - 14). Unlike regular relay outputs,
Bin. out 11 - 14 are bounce-free binary outputs (small signals) and have a
minimal reaction time.
In addition, two high frequency counter inputs for up to 100 kHz are
available for the testing of energy meters. They are described in section
6.4.2, "Counter Inputs 100 kHz (Low Level)" on page 63.
Figure 6-10:
Pin assignment of "ext.
Interf." (upper 16-pole
Lemo socket); view onto
the connector from the
cable wiring side
Table 6-12:
Data of the low-level
binary outputs 11 - 14
Pin
Function
Pin 1
Counter input 1
Pin 2
Counter input 2
Pin 3
Reserved
Pin 4
Neutral (N) connected to GND
Pin 5
Binary output 11
Pin 6
Binary output 12
Pin 7
Binary output 13
Pin 8
Binary output 14
Pin 9
Reserved
Housing
Screen connection
4 Low-Level Transistor Binary Outputs (Bin. out 11 - 14)
Type
Open-collector transistor outputs;
external pull-up resistor
Switching voltage
Max. 15 V
Max. input voltage
±16 V
Switch current
Max. 5 mA (current limited); min. 100 µA
Actualization time
100 µs
Rise time
< 3 µs (V
extern
= 5 V, R
pullup
= 4.7 k
Ω
)
Connection
Connector "ext. Interf." (
CMC 356
rear side)
Insulation
Reinforced insulation to all other potential
groups of the test equipment. GND is
connected to protective earth (PE).
Summary of Contents for CMC 356
Page 1: ...CMC 356 Reference Manual ...
Page 40: ...CMC 356 Reference Manual 40 ...
Page 104: ...CMC 356 Reference Manual 104 ...
Page 108: ...Open Source License Information 108 ...
Page 110: ...Support 110 ...