6.11.7.2
Accuracy of Amplitude Measurement
The measurement values are given as effective values (rms).
The permitted input frequency range for the fundamental wave depends on
the specified sampling rate:
Table 6-47:
Sampling rate and input
frequency range
Valid for fundamental and harmonic waves in specified frequency range; accuracy
refers to full scale.
Table 6-48:
Sampling rate 28.44 kHz,
measurement range
600 V, 100 V, 10 V, 1 V
Table 6-49:
Sampling rate 28.44 kHz,
measurement range
100 mV
Table 6-50:
Sampling rate
9.48 kHz
3.16 kHz,
measurement range
600 V, 100 V, 10 V, 1 V
Sampling rate
Input frequency range
28.44 kHz
100 Hz (= fmin) ... 3200 Hz
9.48 kHz
30 Hz (= fmin) ... 1000 Hz
3.6 kHz
10 Hz (= fmin) ... 350 Hz
Frequency range
Accuracy
Typical
Guaranteed
fmin ... 1 kHz
± 0.1 %
± 0.3 %
fmin ... 10 kHz
+ 0.1 % / - 0.7 %
± 1.1 %
Frequency range
Accuracy
Typical
Guaranteed
fmin ... 1 kHz
± 0.2 %
± 0.5 %
fmin ... 10 kHz
+ 0.2 % / - 1.0 %
± 2.0 %
Frequency range
Accuracy
Typical
Guaranteed
fmin ... 100 Hz
± 0.1 %
± 0.3 %
fmin ... 1 kHz
+ 0.1 % / - 0.5 %
± 0.8 %
fmin ... 4 kHz
(sampling rate = 9.48 kHz)
+ 0.1 % / - 0.8 %
±1.2 %
fmin ... 1.4 kHz
(sampling rate = 3.16 kHz)
+ 0.1 % / - 0.8 %
±1.2 %